DIMENSION? DIMENSION? EXL WITH LM SMN 10472176

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-02-07 for DIMENSION? DIMENSION? EXL WITH LM SMN 10472176 manufactured by Siemens Healthcare Diagnostics Inc - Glasgow.

Event Text Entries

[66681551] The customer contacted the siemens customer care center (ccc). The cause of the discordant amm results is unknown. The issue was resolved by replacement of the flex reagent cartridge with an alternate amm flex reagent cartridge lot and repeat testing. Siemens is investigating the incident.
Patient Sequence No: 1, Text Type: N, H10


[66681552] Discordant low ammonia (amm) results were obtained on a patient sample on the dimension exl system. The patient result was not reported to the physician. The sample was rerun on the same sample on the same instrument after a flex reagent cartridge change and a higher result was obtained and reported. There are no reports of patient intervention or adverse health consequences as a result of the discordant amm results.
Patient Sequence No: 1, Text Type: D, B5


[73363322] Original mdr 2517506-2017-00135 was filed 2/7/2017. Siemens healthcare diagnostic headquarters support center evaluated the incident and concluded their investigation. The cause of the discordant low ammonia results is unknown. There was only one patient sample affected. The data logs from the instrument were analyzed. The low results were processed using flex lot fa7096. There was no indication in the data logs of reagent delivery, foaming, or hemolysis interference based on result monitors for the tests processed using either lot fa7096 or the flex lot fa7123 used for the repeat tests. No systemic instrument or reagent malfunction was confirmed. The instrument is performing within specifications. No further evaluation of the device is required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2517506-2017-00135
MDR Report Key6307017
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-02-07
Date of Report2017-03-24
Date of Event2017-01-10
Date Mfgr Received2017-03-02
Device Manufacturer Date2014-04-11
Date Added to Maude2017-02-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAMES MORGERA
Manufacturer StreetGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE PO BOX 6101
Manufacturer CityNEWARK DE 197146101
Manufacturer CountryUS
Manufacturer Postal197146101
Manufacturer Phone3026318356
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC- BROOKFIELD
Manufacturer StreetREGISTRATION NUMBER 1226181 101 SILVERMINE ROAD
Manufacturer CityBROOKFIELD CT 06804
Manufacturer CountryUS
Manufacturer Postal Code06804
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameDIMENSION?
Generic NameDIMENSION? EXL WITH LM INTEGRATED CHEMISTRY SYSTEM
Product CodeJIF
Date Received2017-02-07
Model NumberDIMENSION? EXL WITH LM
Catalog NumberSMN 10472176
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No0
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Manufacturer AddressGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE PO BOX 6101 NEWARK DE 197146101 US 197146101

Device Sequence Number: 1

Brand NameDIMENSION?
Generic NameDIMENSION? EXL WITH LM INTEGRATED CHEMISTRY SYSTEM
Product CodeJJE
Date Received2017-02-07
Model NumberDIMENSION? EXL WITH LM
Catalog NumberSMN 10472176
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Manufacturer AddressGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE PO BOX 6101 NEWARK DE 197146101 US 197146101


Patients

Patient NumberTreatmentOutcomeDate
10 2017-02-07

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.