JAMSHIDI NEEDLE BONE MARROW 11GX4 ASP DJ4011X

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-02-07 for JAMSHIDI NEEDLE BONE MARROW 11GX4 ASP DJ4011X manufactured by Carefusion, Inc.

Event Text Entries

[67134272] (b)(4) initial emdr submission. A follow up emdr will be submitted upon completion of investigation. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[67134273] Sales rep reported via email: the handle broke off the needle while the needle was in the patient's pelvis. We were able to get the needle out with vice grips and no harm was done to the patient.
Patient Sequence No: 1, Text Type: D, B5


[73687350] (b)(4) follow up emdr for device evaluation. One (1) sample from lot #0000987427 was received for evaluation. During visual analysis failure mode could be confirmed since the handle was detached from the needle. A review of the internal manufacturing device record and raw material history files for the reported lot number was performed and no recorded quality problems or rejections related to this incident were found. It was confirmed that procedural and functional requirements needed for its release were met. ? Lot#0000987427 was manufactured on 01-sep-2016. Defective component: most probable cause of failure mode reported could be related with material provided from (b)(4). A scar (supplier corrective action request) was sent to the supplier ((b)(4)). This failure mode will be entered into the complaint tracking system and tracked & trended for future occurrences of any similar failure modes.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9680904-2017-00099
MDR Report Key6307076
Date Received2017-02-07
Date of Report2017-04-05
Date of Event2017-01-27
Date Mfgr Received2017-01-27
Date Added to Maude2017-02-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactANNA WEHRHEIM
Manufacturer Street75 NORTH FAIRWAY DRIVE
Manufacturer CityVERNON HILLS IL 60061
Manufacturer CountryUS
Manufacturer Postal60061
Manufacturer G1CAREFUSION, INC
Manufacturer StreetZONA FRANCA LAS AMERICAS
Manufacturer CitySANTO DOMINGO
Manufacturer CountryDR
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameJAMSHIDI NEEDLE BONE MARROW 11GX4 ASP
Generic NameTRAY, SURGICAL, NEEDLE
Product CodeFSH
Date Received2017-02-07
Returned To Mfg2017-02-03
Model NumberDJ4011X
Lot Number0000987427
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCAREFUSION, INC
Manufacturer Address75 NORTH FAIRWAY DRIVE VERNON HILLS IL 60061 US 60061


Patients

Patient NumberTreatmentOutcomeDate
10 2017-02-07

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.