MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-02-07 for RELIANCE ENDOSCOPE PROCESSING SYSTEM manufactured by Steris Canada Corporation.
[67208556]
A steris service technician arrived on-site to inspect the reliance endoscope processing system and found that the chemical pump in the processing system had shorted due to a blown fuse. The technician repaired the processor, ran a test cycle, and found the unit to be operational. No additional issues have been reported.
Patient Sequence No: 1, Text Type: N, H10
[67208557]
The user facility reported their reliance endoscope processing system was not operating properly. No report of injury, procedure delays or cancellations.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 9680353-2017-00013 |
MDR Report Key | 6309119 |
Date Received | 2017-02-07 |
Date of Report | 2017-02-07 |
Date of Event | 2017-01-08 |
Date Mfgr Received | 2017-01-08 |
Date Added to Maude | 2017-02-07 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS. KATHRYN CADORETTE |
Manufacturer Street | 5960 HEISLEY ROAD |
Manufacturer City | MENTOR OH 44060 |
Manufacturer Country | US |
Manufacturer Postal | 44060 |
Manufacturer Phone | 4403927231 |
Manufacturer G1 | STERIS CANADA CORPORATION |
Manufacturer Street | 490, ARMAND-PARIS |
Manufacturer City | QUEBEC, QUEBEC GIC 8A3 |
Manufacturer Country | CA |
Manufacturer Postal Code | GIC 8A3 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 0 |
Brand Name | RELIANCE ENDOSCOPE PROCESSING SYSTEM |
Generic Name | RELIANCE EPS |
Product Code | NZA |
Date Received | 2017-02-07 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | STERIS CANADA CORPORATION |
Manufacturer Address | 490 ARMAND-PARIS QUEBEC, QUEBEC GIC 8A3 CA GIC 8A3 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-02-07 |