BN II SYSTEM SMN 10454733

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-02-08 for BN II SYSTEM SMN 10454733 manufactured by Siemens Healthcare Diagnostics Products Gmbh.

Event Text Entries

[67250141] The cause of the imprecise replicate iggc results is unknown. A siemens field service engineer was dispatched to the site and was not able to reproduce the imprecision. Instrument probes were replaced proactively and precision studies were acceptable. The instrument is performing within specifications. No further evaluation of the device is required.
Patient Sequence No: 1, Text Type: N, H10


[67250142] Imprecise replicate immunoglobulin g (iggc) results were obtained on a patient cerebrospinal fluid (csf) sample on the bnii system. It is unknown if results were reported to the physician. There are no known reports of patient intervention or adverse health consequences due to the imprecise replicate iggc results.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610806-2017-00020
MDR Report Key6312825
Date Received2017-02-08
Date of Report2017-02-08
Date of Event2016-12-14
Date Mfgr Received2017-01-09
Date Added to Maude2017-02-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAMES MORGERA
Manufacturer StreetGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE PO BOX 6101
Manufacturer CityNEWARK DE 197146101
Manufacturer CountryUS
Manufacturer Postal197146101
Manufacturer Phone3026318356
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer StreetEMIL VON BEHRING STRASSE 76
Manufacturer CityMARBURG, D-35041
Manufacturer CountryGM
Manufacturer Postal CodeD-35041
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameBN II SYSTEM
Generic NameBN II SYSTEM
Product CodeDEW
Date Received2017-02-08
Model NumberBN II SYSTEM
Catalog NumberSMN 10454733
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer AddressEMIL VON BEHRING STRASSE 76 MARBURG, D-35041 GM D-35041


Patients

Patient NumberTreatmentOutcomeDate
10 2017-02-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.