PANOVIEW TELESCOPE 8974.402

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2017-02-10 for PANOVIEW TELESCOPE 8974.402 manufactured by Richard Wolf Gmbh (germany).

Event Text Entries

[67439333] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[67439334] During hysteroresectoscopy of submucous node, at the third attempt of cutting disappeared the image on the screen. The operation was suspended, hysteroresectoscope derived from the cavity of the uterus: checked connection cables, light guide, light source, and camera - all in production mode; removed internal and external tubes - discovered damaged cutting electrode / loop / , which exploded and carbon deposits on hysteroresectoscope in place of the projection of the cutting electrode , which prevented the appearance of the image. To remove the cutting electrode was not possible because of the so-called "cleavage" at the distal end of the hysteroscope. Were used in the electrocoagulator of the firm "martin", mode cut g2 cut with "gain". Important information: the operation was repeated and the patient is in satisfactory condition. (b)(4) has not received a complaint.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9611102-2017-00002
MDR Report Key6321640
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2017-02-10
Date of Report2017-01-10
Date of Event2016-12-06
Date Mfgr Received2017-01-10
Device Manufacturer Date2016-07-21
Date Added to Maude2017-02-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JOZSEF RADOVITS
Manufacturer Street353 CORPORATE WOODS PARKWAY
Manufacturer CityVERNON HILLS IL 600613110
Manufacturer CountryUS
Manufacturer Postal600613110
Manufacturer Phone8479131113
Manufacturer G1RICHARD WOLF GMBH (GERMANY)
Manufacturer StreetREGISTRATION NUMBER 9611102 32 PFORZHEIMER STREET
Manufacturer CityKNITTLINGEN, GERMANY 75438
Manufacturer CountryGM
Manufacturer Postal Code75438
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePANOVIEW TELESCOPE
Generic NameENDOSCOPE
Product CodeGCM
Date Received2017-02-10
Returned To Mfg2017-01-19
Model Number8974.402
Catalog Number8974.402
OperatorPHYSICIAN
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerRICHARD WOLF GMBH (GERMANY)
Manufacturer AddressREGISTRATION NUMBER 9611102 32 PFORZHEIMER STREET KNITTLINGEN, GERMANY 75438 GM 75438


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-02-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.