[67399310] (b)(4). Rw (b)(4) is the manufacturer of this device.
Patient Sequence No: 1, Text Type: N, H10


[67399311] During hysteroresectoscopy of submucous node, at the third attempt of cutting disappeared the image on the screen. The operation was suspended, hysteroresectoscope derived from the cavity of the uterus: checked connection cables, light guide, light source, and camera - all in production mode; removed internal and external tubes - discovered damaged cutting electrode / loop / , which exploded and carbon deposits on hysteroresectoscope in place of the projection of the cutting electrode , which prevented the appearance of the image. To remove the cutting electrode was not possible because of the so-called "cleavage" at the distal end of the hysteroscope. Were used in the electrocoagulator of the firm "martin", mode cut g2 cut with "gain". Important information: the operation was repeated and the patient is in satisfactory condition. Richard wolf (b)(4) has not received a complaint.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9611102-2017-00003
MDR Report Key6321651
Report SourceCOMPANY REPRESENTATIVE
Date Received2017-02-10
Date of Report2017-01-10
Date of Event2016-12-06
Date Mfgr Received2017-01-10
Device Manufacturer Date2015-07-25
Date Added to Maude2017-02-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JOZSEF RADOVITS
Manufacturer Street353 CORPORATE WOODS PARKWAY
Manufacturer CityVERNON HILLS IL 600613110
Manufacturer CountryUS
Manufacturer Postal600613110
Manufacturer Phone8479131113
Manufacturer G1RICHARD WOLF GMBH (GERMANY)
Manufacturer StreetREGISTRATION NUMBER 9611102 32 PFORZHEIMER STREET
Manufacturer CityKNITTLINGEN, GERMANY 75438
Manufacturer CountryGM
Manufacturer Postal Code75438
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameMONOPOLAR CUTTING ELECTRODE, 21FR
Generic NameELECTRODE
Product CodeJOS
Date Received2017-02-10
Returned To Mfg2017-01-19
Model Number8353.131
Catalog Number8353.131
Lot Number1277723
OperatorPHYSICIAN
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerRICHARD WOLF GMBH (GERMANY)
Manufacturer AddressREGISTRATION NUMBER 9611102 32 PFORZHEIMER STREET KNITTLINGEN, GERMANY 75438 GM 75438


Patients

Patient NumberTreatmentOutcomeDate