ORAL CARE KIT,W/MOIST,PEROXIDE 12260 991096883

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,other,use report with the FDA on 2017-02-10 for ORAL CARE KIT,W/MOIST,PEROXIDE 12260 991096883 manufactured by Halyard Health.

Event Text Entries

[67279273] The sample was received and evaluated. One sample was received with the package label. The sample was received inside a bio-hazard sample bag. The sample was disinfected inside the lab hood using caviwipes. The swab appears in clean condition with no obvious visible signs of use. The swab stick was broken at approximately 3. 25" from the end. Under magnification, stress whitening was observed on the broken ends. Lateral white striations were also observed adjacent to the break. The swab sponge was securely attached to the stick. Root has not been determined and the investigation is ongoing. The device history record for the lot number, m6228a601, involved in this complaint was reviewed and the material was produced according to the manufacturing procedures and met the quality requirements. Halyard health has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred(b)(4).
Patient Sequence No: 1, Text Type: N, H10


[67279274] It was reported that an oral swab broke while in use with a patient. Additional information was received on 24-jan-2017 that states, "while using an oral swab, the stick broke. It snapped in half with little pressure being applied. " additional information was received on 31-jan-2017 that states, the two issues reported on (b)(6) 2017 are the same incident and only one should be reported for (b)(6) 2017. The additional email was provided to give more detailed information and there was only one patient involved. No additional information was provided.
Patient Sequence No: 1, Text Type: D, B5


[71609399] Pro code: -kxf, mfg site for devices: (b)(4). All information reasonably known as of 31mar2017 has been included in this health authority report. Should additional information be obtained, a follow-up health authority report will be provided. The information provided by halyard health represents all of the known information at this time. Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to halyard health. Halyard health has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred. This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8030647-2017-00026
MDR Report Key6322094
Report SourceHEALTH PROFESSIONAL,OTHER,USE
Date Received2017-02-10
Date of Report2017-03-14
Date of Event2017-01-17
Date Mfgr Received2017-03-14
Device Manufacturer Date2016-08-15
Date Added to Maude2017-02-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. LISA CLARK
Manufacturer Street5405 WINDWARD PARKWAY
Manufacturer CityALPHARETTA GA 30004
Manufacturer CountryUS
Manufacturer Postal30004
Manufacturer Phone4704485444
Manufacturer G1AVENT S DE RL DE CV
Manufacturer StreetCARRETERA INTERNACIONAL SALIDA NORTE NO. 1053
Manufacturer CityCP
Manufacturer CountryMX
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameORAL CARE KIT,W/MOIST,PEROXIDE
Generic NameVAP ORAL CARE PRODUCTS
Product CodeKXF
Date Received2017-02-10
Returned To Mfg2017-01-25
Model Number12260
Catalog Number991096883
Lot NumberM6228A601
Device Expiration Date2018-08-15
OperatorOTHER HEALTH CARE PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHALYARD HEALTH
Manufacturer Address5405 WINDWARD PARKWAY ALPHARETTA GA 30004 US 30004


Patients

Patient NumberTreatmentOutcomeDate
10 2017-02-10

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.