SPS-1 STATIC PRESERVATION SOLUTION SPS-1_1L

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-02-10 for SPS-1 STATIC PRESERVATION SOLUTION SPS-1_1L manufactured by Organ Recovery Systems, Inc.

Event Text Entries

[67535845] Investigation of this issue and the recall activities for these lots are in progress. Fda will be notified upon additional information available.
Patient Sequence No: 1, Text Type: N, H10


[67535846] Four complaints were received from customers who observed an uncharacteristic odor emanating from (b)(4) (static preservation solution) solution bags and the solution itself. Related ors(organ recovery systems, inc. ) complaint numbers: (b)(4). Lots involved: sps-1_1l: pbr-0060-392, sps-1_1l: pbr-0060-386, sps-1_2l: pbr-0074-330, sps-1_2l: pbr-0074-337. In cases where the 1l and 2l were included in the complaint, the customers weren't sure which bag the odor was coming from. After initial investigation it is confirmed that the device is malfunctioned. Complaints were determined to be potentially reportable. Ors management decided to initiate a voluntary product recall in parallel with the investigation. The fda (b)(4) was notified of the initial actions for lots pbr-0060-392 and pbr-0074-330 on 15dec16. Communication with the fda continued throughout the investigation as the additional complaints were received and additional actions taken. Date complaint received for lot pbr-0060-386: january 11, 2017. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004068499-2017-00002
MDR Report Key6324278
Date Received2017-02-10
Date of Report2017-02-10
Date of Event2017-01-08
Date Mfgr Received2017-01-11
Device Manufacturer Date2016-05-17
Date Added to Maude2017-02-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. PJ PARDO
Manufacturer StreetONE PIERCE PLACE SUITE 475W
Manufacturer CityITASCA IL 60143
Manufacturer CountryUS
Manufacturer Postal60143
Manufacturer Phone8478242600
Manufacturer G1ORGAN RECOVERY SYSTEMS, INC
Manufacturer StreetONE PIERCE PLACE SUITE 475W
Manufacturer CityITASCA IL 60143
Manufacturer CountryUS
Manufacturer Postal Code60143
Single Use3
Previous Use Code3
Removal Correction NumberZ-1111-2017
Event Type3
Type of Report0

Device Details

Brand NameSPS-1 STATIC PRESERVATION SOLUTION
Generic NameSPS-1 (UW SOLUTION)
Product CodeKDN
Date Received2017-02-10
Model NumberSPS-1_1L
Catalog NumberSPS-1_1L
Lot NumberPBR-0060-386
Device Expiration Date2018-05-17
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No0
Device Event Key0
ManufacturerORGAN RECOVERY SYSTEMS, INC
Manufacturer AddressONE PIERCE PLACE SUITE 475W ITASCA IL 60143 US 60143


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-02-10

© 2026 FDA.report