SEDLINE MONITOR 16505

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-02-13 for SEDLINE MONITOR 16505 manufactured by Masimo - 40 Parker.

Event Text Entries

[67249784] The returned device was evaluated. During inspection, the unit was found to have a cracked housing. The device was able to power on and obtain valid readings with a sedline tester; however the readings could not be seen as the display was dim. The backlight inverter was found to be defective. The display was temporarily replaced with a known good display and the values verified as correct. The backlight inverter was replaced and the unit functioned as designed.
Patient Sequence No: 1, Text Type: N, H10


[67249885] The customer reported the device will not stay powered on. No consequences or impact to patient were reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2031172-2017-00194
MDR Report Key6325664
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2017-02-13
Date of Event2017-01-22
Date Mfgr Received2017-01-22
Device Manufacturer Date2017-01-31
Date Added to Maude2017-02-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCHARLENE JOHNSON
Manufacturer Street52 DISCOVERY
Manufacturer CityIRVINE CA 926181604
Manufacturer CountryUS
Manufacturer Postal926181604
Manufacturer Phone9492977000
Manufacturer G1INDUSTRIAL VALLERA DE MEXICALI S.A DE C.V.
Manufacturer StreetCALZADA DEL ORO, NO. 2001 PARQUE INDUSTRIAL PALACO
Manufacturer CityMEXICALI, BAJA CALIFORNIA 21600
Manufacturer CountryMX
Manufacturer Postal Code21600
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSEDLINE MONITOR
Generic NameELECTROENCEPHALOGRAPH
Product CodeGWQ
Date Received2017-02-13
Returned To Mfg2017-01-27
Model Number16505
Catalog Number16505
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMASIMO - 40 PARKER
Manufacturer Address40 PARKER IRVINE CA 926181604 US 926181604


Patients

Patient NumberTreatmentOutcomeDate
10 2017-02-13

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.