ELECSYS MYOGLOBIN IMMUNOASSAY 12178214160

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-02-13 for ELECSYS MYOGLOBIN IMMUNOASSAY 12178214160 manufactured by Roche Diagnostics.

Event Text Entries

[67339259] Expiration date was provided as (b)(6) 2017. This event occurred in (b)(6). (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[67339260] The customer questioned high results for 1 patient tested for elecsys myoglobin immunoassay (myoglobin) on a cobas 8000 e 602 module that did not fit the clinical picture for this patient. The patient was hospitalized due to kidney failure. The patient? S myoglobin results for more than 2 weeks were >3000 ug/l. The customer inquired if there is an interference with the myoglobin assay and the medications the patient is being treated with. Refer to attached data for the patient results. No adverse event occurred. The e602 module serial number was (b)(4). A specific root cause could not be identified for this event. Additional information was requested for investigation but was not provided. Since the patient has kidney failure, myoglobin results could be elevated due to the disease. It is also possible that the medications the patient is being treated with are having a cardio toxic effect increasing the myoglobin results. However, since no patient samples could be provided, this could not be confirmed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2017-00299
MDR Report Key6326819
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-02-13
Date of Report2017-02-13
Date of Event2016-12-31
Date Mfgr Received2017-01-25
Date Added to Maude2017-02-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameELECSYS MYOGLOBIN IMMUNOASSAY
Generic NameMYOGLOBIN IMMUNOLOGICAL TEST SYSTEM
Product CodeDDR
Date Received2017-02-13
Model NumberNA
Catalog Number12178214160
Lot Number133148-01
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2017-02-13