CART 9734056 S7 STAFF SHRT 100-120V INTL

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-02-13 for CART 9734056 S7 STAFF SHRT 100-120V INTL manufactured by Medtronic Navigation, Inc. (louisville).

Event Text Entries

[67584981] No patient information provided as no patient was involved in this concern. The initial report occurred on (b)(6) 2016 and was found to be a non-reportable malfunction based on the information available at the time the event occurred. On 01/17/2017, a retrospective review related to a capa was completed for all events where the reported malfunction was not previously associated with serious injuries; the reporting determination has been revised to consider these malfunction events reportable. The device was returned to the manufacturer for analysis. The device was found to be fully functional with no problem found. The reported event could not be duplicated by medtronic personnel.
Patient Sequence No: 1, Text Type: N, H10


[67584982] A medtronic representative reported that the scu is cycling power. There was no patient present when this issue was identified.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1723170-2017-00650
MDR Report Key6327594
Date Received2017-02-13
Date of Report2017-02-13
Date of Event2016-11-18
Date Mfgr Received2017-01-17
Device Manufacturer Date2016-06-17
Date Added to Maude2017-02-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactROB EDWARD
Manufacturer Street826 COAL CREEK CIRCLE
Manufacturer CityLOUISVILLE CO 80027
Manufacturer CountryUS
Manufacturer Postal80027
Manufacturer Phone7208902154
Manufacturer G1MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
Manufacturer Street826 COAL CREEK CIRCLE
Manufacturer CityLOUISVILLE CO 80027
Manufacturer CountryUS
Manufacturer Postal Code80027
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCART 9734056 S7 STAFF SHRT 100-120V INTL
Generic NameSEE H10)
Product CodeBZN
Date Received2017-02-13
Returned To Mfg2016-12-08
Model NumberS7
Catalog Number9734056
OperatorMEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC NAVIGATION, INC. (LOUISVILLE)
Manufacturer Address826 COAL CREEK CIRCLE LOUISVILLE CO 80027 US 80027


Patients

Patient NumberTreatmentOutcomeDate
10 2017-02-13

© 2026 FDA.report