VISIONPRO 26" LED DISPLAY INTERNATIONAL KIT 0240031020I

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-02-14 for VISIONPRO 26" LED DISPLAY INTERNATIONAL KIT 0240031020I manufactured by Stryker Endoscopy-san Jose.

Event Text Entries

[67592696] Additional information will be provided once the investigation has been completed. The device manufacturer date is not known at this time. However, should it become available it will be provided in future reports. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[67592697] It was reported that there was a loss of image during the procedure. There was a 20 minute delay while the patient received additional anesthesia. The procedure was completed successfully.
Patient Sequence No: 1, Text Type: D, B5


[80447157] The product was not returned for investigation therefore the reported failure mode was not confirmed. The reported failure mode will be monitored for future reoccurrence. Alleged failure: loss of image. Probable root cause: tx main/ antenna/front board, display main/antenna board/silex module, connectors, display firmware/software, transmitter software/firmware, available wireless channels, display power supply, token / token slot, wireless interference, use error. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[80447158] It was reported that there was a loss of image during the procedure. There was a 20 minute delay while the patient received additional anesthesia. The procedure was completed successfully.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0002936485-2017-00171
MDR Report Key6328920
Date Received2017-02-14
Date of Report2017-06-05
Date of Event2017-01-20
Date Mfgr Received2017-01-20
Date Added to Maude2017-02-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. VICTORIA MILICH
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal95138
Manufacturer Phone4087542000
Manufacturer G1STRYKER ENDOSCOPY-SAN JOSE
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal Code95138
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVISIONPRO 26" LED DISPLAY INTERNATIONAL KIT
Generic NameCAMERA, TELEVISION, SURGICAL, WITHOUT AUDIO
Product CodeFWB
Date Received2017-02-14
Catalog Number0240031020I
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER ENDOSCOPY-SAN JOSE
Manufacturer Address5900 OPTICAL COURT SAN JOSE CA 95138 US 95138


Patients

Patient NumberTreatmentOutcomeDate
10 2017-02-14

© 2024 FDA.report