VISERA ELITE XENON LIGHT SOURCE CLV-S190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-02-15 for VISERA ELITE XENON LIGHT SOURCE CLV-S190 manufactured by Olympus Medical Systems Corp..

Event Text Entries

[67708209] The subject circuit board of the clv-s190 was returned to olympus for evaluation. Olympus replaced the circuit board of the spare clv-s190 with the subject circuit board and confirmed the spare clv-s190. The spare clv-s190 worked normally with the subject circuit board and there was no other abnormality found. There was no similar medical device report of clv-s190 in the past. There were no further details provided at this time. If significant additional information is received, this report will be supplemented.
Patient Sequence No: 1, Text Type: N, H10


[67708210] During the prostate plasma resection with the subject clv-s190, when the resectoscope was moved, the image of the monitor flashed. The user facility replaced the subject clv-s190 with the unspecified similar device to complete the procedure. There was no report of the patient injury other than replacing the device. When the field service engineer of olympus checked the subject clv-s190 at the branch office, the phenomenon was reproduced. The circuit board of the subject clv-s190 was replaced, and the circuit board was returned to olympus for evaluation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8010047-2017-00173
MDR Report Key6331150
Date Received2017-02-15
Date of Event2016-11-04
Date Mfgr Received2016-11-07
Date Added to Maude2017-02-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR HIROKI MORIYAMA
Manufacturer Street2951 ISHIKAWA-CHO
Manufacturer CityHACHIOJI-SHI, TOKYO-TO 192-8507
Manufacturer CountryJA
Manufacturer Postal192-8507
Manufacturer Phone8142642517
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameVISERA ELITE XENON LIGHT SOURCE
Generic NameXENON LIGHT SOURCE
Product CodeGCT
Date Received2017-02-15
Returned To Mfg2017-01-06
Model NumberCLV-S190
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS MEDICAL SYSTEMS CORP.
Manufacturer Address2951 ISHIKAWA-CHO HACHIOJI-SHI, TOKYO-TO 192-8507 JA 192-8507


Patients

Patient NumberTreatmentOutcomeDate
10 2017-02-15

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.