BN PROSPEC ANALYZER SMN 10461865

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-02-15 for BN PROSPEC ANALYZER SMN 10461865 manufactured by Siemens Healthcare Diagnostics Products Gmbh.

Event Text Entries

[67500449] The cause of the discordant low csf albumin results is an incorrect probe alignment. A siemens healthcare field service engineer was dispatched to the account. During instrument maintenance an incorrect probe alignment was identified and immediately corrected. After the bn prospec maintenance was performed there were no further discrepant results observed by the customer. Siemens headquarters support center representative evaluated the quality control data from the time of the incident and noted that qc had been within laboratory ranges. The instrument is performing within specifications. No further evaluation of the device is required.
Patient Sequence No: 1, Text Type: N, H10


[67500450] Discordant low albumin results were obtained on cerebrospinal fluid (csf) patient samples on the bn prospec instrument. Results were reported to the physician(s). The same samples were repeated and higher results were obtained. Corrected reports were issued. There are no reports of patient intervention or adverse health consequences as a result of the discordant low albumin results.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610806-2017-00022
MDR Report Key6332103
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-02-15
Date of Report2017-02-15
Date of Event2017-01-11
Date Mfgr Received2017-01-16
Date Added to Maude2017-02-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAMES MORGERA
Manufacturer StreetGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE PO BOX 6101
Manufacturer CityNEWARK DE 197146101
Manufacturer CountryUS
Manufacturer Postal197146101
Manufacturer Phone3026318356
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer StreetEMIL VON BEHRING STRASSE 76
Manufacturer CityMARBURG, D-35041
Manufacturer CountryGM
Manufacturer Postal CodeD-35041
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBN PROSPEC ANALYZER
Generic NameBN PROSPEC ANALYZER
Product CodeJZW
Date Received2017-02-15
Catalog NumberSMN 10461865
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer AddressEMIL VON BEHRING STRASSE 76 MARBURG, D-35041 GM D-35041


Patients

Patient NumberTreatmentOutcomeDate
10 2017-02-15

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.