ILED ILED3K 1537478

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-02-17 for ILED ILED3K 1537478 manufactured by Trumpf Medizin Systeme Gmbh + Co. Kg.

Event Text Entries

[68012981] The affective lamp body was replaced. The defective parts were returned to trumpf medical for examination.
Patient Sequence No: 1, Text Type: N, H10


[68012982] During preventative maintenance of the iled 3k, peeling paint was discovered on the unit's lamp head. No injuries were reported.
Patient Sequence No: 1, Text Type: D, B5


[74536021] The affected device was replaced at the facility. The returned, defective parts were investigated and evaluated. Our investigation has identified steps in the manufacturing process which combined with non-approved cleaners can lead to the paint chipping. The damage to the painted surfaces usually does not develop suddenly, but develops over time. The instructions for use state that the devices must be checked for proper condition prior to each use. Damaged devices must not be used. If damage to the surface coating is recognized and removed, there is no danger to the patient. The surface coating process will be updated to include additional testing and verification.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9681407-2017-00007
MDR Report Key6339587
Date Received2017-02-17
Date of Report2017-01-19
Date of Event2017-01-19
Device Manufacturer Date2010-10-12
Date Added to Maude2017-02-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationSERVICE PERSONNEL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. STEFFEN ULBRICH
Manufacturer StreetCARL-ZEISS STRASSE 7-9
Manufacturer CitySAALFELD, 07318
Manufacturer CountryGM
Manufacturer Postal07318
Manufacturer Phone6715864140
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameILED
Generic NameILED 3 LAMP, OPERATING-ROOM
Product CodeFQP
Date Received2017-02-17
Model NumberILED3K
Catalog Number1537478
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTRUMPF MEDIZIN SYSTEME GMBH + CO. KG
Manufacturer AddressCARL-ZEISS STRASSE 7-9 SAALFELD, 07318 GM 07318


Patients

Patient NumberTreatmentOutcomeDate
10 2017-02-17

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.