MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-02-17 for STATIC AND DYNAMIC COMPLIANCE VMAX ENCORE 777401-101 manufactured by Carefusion, Inc.
[68014750]
(b)(4). The carefusion technical support assisted the customer with performing firmware upgrade which solved the stopping issue. Due to the event being resolved by upgrading the firmware; no parts will be sent back to carefusion.
Patient Sequence No: 1, Text Type: N, H10
[68014751]
The customer reported while using the vmax encore device, the treadmill would stop during use. The issue occurred during patient use but no patient injury was reported.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2021710-2017-05437 |
MDR Report Key | 6340693 |
Date Received | 2017-02-17 |
Date of Report | 2017-02-17 |
Date of Event | 2017-02-06 |
Date Mfgr Received | 2017-02-06 |
Device Manufacturer Date | 2011-09-01 |
Date Added to Maude | 2017-02-17 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | JILL RITTORNO |
Manufacturer Street | 22745 SAVI RANCH PARKWAY |
Manufacturer City | YORBA LINDA CA 92887 |
Manufacturer Country | US |
Manufacturer Postal | 92887 |
Manufacturer G1 | CAREFUSION, INC |
Manufacturer Street | 1100 BIRD CENTER DR. |
Manufacturer City | PALM SPRINGS CA 92262 |
Manufacturer Country | US |
Manufacturer Postal Code | 92262 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 0 |
Brand Name | STATIC AND DYNAMIC COMPLIANCE |
Generic Name | CALCULATOR, PREDICTED VALUES, PULMONARY FUNCTION |
Product Code | BTY |
Date Received | 2017-02-17 |
Model Number | VMAX ENCORE |
Catalog Number | 777401-101 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | * |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | CAREFUSION, INC |
Manufacturer Address | 22745 SAVI RANCH PARKWAY YORBA LINDA CA 92887 US 92887 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2017-02-17 |