TOOTHETTE? ORAL CARE UNTREATED DISPOSABLE ORAL SWAB 6070-X

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-02-17 for TOOTHETTE? ORAL CARE UNTREATED DISPOSABLE ORAL SWAB 6070-X manufactured by Sage Products Llc.

Event Text Entries

[68087930] Reporter discarded the involved product and did not provide photographs, however, provided lot information. Production history records for the reported lot number were reviewed and all in-process checks indicated passing results. No anomalies were identified during production of the product. The root cause could not be determined. Potential root causes could be insufficient amount of glue delivered or inadequate glue adhesion during the production of the device. (b)(4); method: no testing methods performed; results: no results available since no evaluation performed; conclusion: device discarded by user, unable to follow-up. Discarded by user.
Patient Sequence No: 1, Text Type: N, H10


[68087931] Report received of a malfunction resulting in an applicator swab disengagement. On an unknown date, oral care was performed on a female patient. The reporter stated the green foam disengaged from the stick. The reporter stated the foam was successfully removed from the patient's oral cavity and no injury occurred. The reporter stated the patient did not bite on the device. The device was discarded and no pictures were available. Although requested, no additional information was available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1419181-2017-00001
MDR Report Key6341103
Date Received2017-02-17
Date of Report2017-02-15
Date Mfgr Received2017-01-18
Device Manufacturer Date2015-09-27
Date Added to Maude2017-02-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. HEATHER SIRIANNI
Manufacturer Street3909 THREE OAKS ROAD
Manufacturer CityCARY IL 60013
Manufacturer CountryUS
Manufacturer Postal60013
Manufacturer Phone8154554700
Manufacturer G1SAGE PRODUCTS LLC
Manufacturer Street3909 THREE OAKS ROAD
Manufacturer CityCARY IL 60013
Manufacturer CountryUS
Manufacturer Postal Code60013
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameTOOTHETTE? ORAL CARE UNTREATED DISPOSABLE ORAL SWAB
Generic NameAPPLICATOR SWAB
Product CodeKXF
Date Received2017-02-17
Model Number6070-X
Catalog Number6070-X
Lot Number51557
Device Expiration Date2017-09-26
OperatorPATIENT FAMILY MEMBER OR FRIEND
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSAGE PRODUCTS LLC
Manufacturer Address3909 THREE OAKS ROAD CARY IL 60013 US 60013


Patients

Patient NumberTreatmentOutcomeDate
10 2017-02-17

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.