RANEY CLIP N/A FFP013P

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-02-21 for RANEY CLIP N/A FFP013P manufactured by Aesculap, Inc..

Event Text Entries

[67860001]
Patient Sequence No: 1, Text Type: N, H10


[67860002] We opened a pack of raney clips (lot#52274399) and determined the clips would not appropriately fire with any of the 4 clip appliers. We opened a different package (lot#52153841) and all 4 raney clip appliers functioned appropriately. I then grabbed another 2 packages of raney clips, one with a completely different lot # (52187385), and one with the same "bad" lot (52274399). The lot# 52187385 fired correctly with every clip applier. The lot#52274399 did not fire correctly with any of the 4 clip appliers. It was determined that the issue of misfire was not because of broken clip appliers, but of a bad lot# with the clips. Lead and manager of service line informed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number6345873
MDR Report Key6345873
Date Received2017-02-21
Date of Report2017-01-18
Date of Event2017-01-12
Report Date2017-01-18
Date Reported to FDA2017-01-18
Date Reported to Mfgr2017-01-18
Date Added to Maude2017-02-21
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRANEY CLIP
Generic NameCLIP, SCALP
Product CodeHBO
Date Received2017-02-21
Model NumberN/A
Catalog NumberFFP013P
Lot Number52274399
Device AvailabilityY
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerAESCULAP, INC.
Manufacturer Address3773 CORPORATE PKWY. CENTER VALLEY PA 18034 US 18034


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-02-21

© 2026 FDA.report