DADE ACTIN FSL ACTIVATED PTT REAGENT B4219-2

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2005-09-08 for DADE ACTIN FSL ACTIVATED PTT REAGENT B4219-2 manufactured by Dade Behring Gmbh.

Event Text Entries

[387235] Erroneous short ptt reuslts of <21 seconds were obtained on the sysmex ca-1500 system. The results were reported to the physician despite the error flags due to user error. Pt was on heparin therapy before the erroneous results were reported but it is unknown if the treatment was altered due to the erroneous results. The pt died. Analysis of instrument data indicates possible sample integrity issue. No additional info could be obtained from the hosp contact.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610806-2005-00020
MDR Report Key634634
Report Source05
Date Received2005-09-08
Date of Report2005-09-07
Date of Event2005-08-24
Date Mfgr Received2005-08-25
Device Manufacturer Date2005-03-01
Date Added to Maude2005-09-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactVICTOR CARRIO
Manufacturer StreetP.O. BOX 6101, MS 514
Manufacturer CityNEWARK DE 197146101
Manufacturer CountryUS
Manufacturer Postal197146101
Manufacturer Phone3026317626
Manufacturer G1DADE BEHRING MARBURG GMBH
Manufacturer Street76 EMIL VON BEHRING STRASSE
Manufacturer CityMARBURG
Manufacturer CountryGM
Single Use3
Remedial ActionOT
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameDADE ACTIN FSL ACTIVATED PTT REAGENT
Generic NamePARTIAL THROMBOPLASTIN TIME REAGENT
Product CodeGIT
Date Received2005-09-08
Model NumberNA
Catalog NumberB4219-2
Lot Number527336A
ID Number*
Device Expiration Date2007-02-21
Device AvailabilityN
Device AgeNA
Device Eval'ed by MfgrN
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key624203
ManufacturerDADE BEHRING GMBH
Manufacturer Address* MARBURG GM
Baseline Brand NameDADE ACTIN FSL ACTIVATED PTT REAGENT
Baseline Generic NamePARTIAL THROMBOPLASTIN TIME REAGENT
Baseline Model NoNA
Baseline Catalog NoB4219-2
Baseline ID*


Patients

Patient NumberTreatmentOutcomeDate
101. Death 2005-09-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.