[68223639] The device was returned dirty and corroded. The cannula of the scissor was cracked and rusted. The malfunction was due to improper cleaning and maintenance. There was no impact to the patient.
Patient Sequence No: 1, Text Type: N, H10


[68223900] During surgery the surgeon noted hearing a snapping sound coming from the scissors. The surgeon discontinued use of the device and there was no impact to the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3019924-2017-00003
MDR Report Key6347811
Date Received2017-02-21
Date of Report2017-02-20
Date of Event2017-01-20
Date Mfgr Received2017-01-20
Device Manufacturer Date2014-10-27
Date Added to Maude2017-02-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JOHN RAY
Manufacturer Street8415 154TH AVE NE
Manufacturer CityREDMOND WA 98052
Manufacturer CountryUS
Manufacturer Postal98052
Manufacturer Phone4255560544
Manufacturer G1MICROSURGICAL TECHNOLOGY INC
Manufacturer Street8415 154TH AVE NE
Manufacturer CityREDMOND WA 98052
Manufacturer CountryUS
Manufacturer Postal Code98052
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NamePACKER CHANG IOL CUTTER
Generic NameOPHTHALMIC SCISSOR
Product CodeHNF
Date Received2017-02-21
Returned To Mfg2017-02-14
Model NumberDFH-0012
Catalog NumberDFH-0012
Lot Number064410
OperatorPHYSICIAN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMICROSURGICAL TECHNOLOGY INC
Manufacturer Address8415 154TH AVE NE REDMOND WA 98052 US 98052


Patients

Patient NumberTreatmentOutcomeDate
10 2017-02-21

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.