MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-02-27 for MDRC-1119 WITH PROTECTIVE COVER KIT MDRC-1119 PROTECTIVE COVER KIT C9826109 manufactured by Barco Nv.
[68768623]
The display nor its front cover were returned to the manufacturer. However based on photographs and other information from the customer and in-house testing it was concluded that the metal strips that are applied (glued) to the display for fixation of the cover, were clearly re-used. As a result the strips are not flat but bended, resulting in a decreased contact surface with the front cover. It is obviously not the intention to re-use the fixation strips but this was not specifically indicated in the installation instructions. A specific warning will be added in the installation instructions.
Patient Sequence No: 1, Text Type: N, H10
[68768624]
During usage of the system, the front cover detached from the display and hit the patient on the cheek. A moderate injury was reported.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 9613188-2017-00002 |
MDR Report Key | 6359988 |
Date Received | 2017-02-27 |
Date of Report | 2017-02-02 |
Date Mfgr Received | 2016-12-21 |
Date Added to Maude | 2017-02-27 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 0 |
Initial Report to FDA | 0 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MR LIEVEN DE WANDEL |
Manufacturer Street | BENELUXPARK 21 |
Manufacturer City | KORTRIJK, 8500 |
Manufacturer Country | BE |
Manufacturer Postal | 8500 |
Manufacturer Phone | 56 233459 |
Manufacturer G1 | BARCO |
Manufacturer Street | PRESIDENT KENNEDYPARK 35 |
Manufacturer City | KORTRIJK, 8500 |
Manufacturer Country | BE |
Manufacturer Postal Code | 8500 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 0 |
Brand Name | MDRC-1119 WITH PROTECTIVE COVER KIT |
Generic Name | CLINICAL REVIEW DISPLAY |
Product Code | PGY |
Date Received | 2017-02-27 |
Model Number | MDRC-1119 PROTECTIVE COVER KIT |
Catalog Number | C9826109 |
Lot Number | N/A |
Device Availability | N |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | BARCO NV |
Manufacturer Address | BENELUXPARK 21 KORTRIJK, 8500 BE 8500 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Congenital Not Applicablenomaly | 2017-02-27 |