MDRC-1119 WITH PROTECTIVE COVER KIT MDRC-1119 PROTECTIVE COVER KIT C9826109

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-02-27 for MDRC-1119 WITH PROTECTIVE COVER KIT MDRC-1119 PROTECTIVE COVER KIT C9826109 manufactured by Barco Nv.

Event Text Entries

[68768623] The display nor its front cover were returned to the manufacturer. However based on photographs and other information from the customer and in-house testing it was concluded that the metal strips that are applied (glued) to the display for fixation of the cover, were clearly re-used. As a result the strips are not flat but bended, resulting in a decreased contact surface with the front cover. It is obviously not the intention to re-use the fixation strips but this was not specifically indicated in the installation instructions. A specific warning will be added in the installation instructions.
Patient Sequence No: 1, Text Type: N, H10


[68768624] During usage of the system, the front cover detached from the display and hit the patient on the cheek. A moderate injury was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9613188-2017-00002
MDR Report Key6359988
Date Received2017-02-27
Date of Report2017-02-02
Date Mfgr Received2016-12-21
Date Added to Maude2017-02-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactMR LIEVEN DE WANDEL
Manufacturer StreetBENELUXPARK 21
Manufacturer CityKORTRIJK, 8500
Manufacturer CountryBE
Manufacturer Postal8500
Manufacturer Phone56 233459
Manufacturer G1BARCO
Manufacturer StreetPRESIDENT KENNEDYPARK 35
Manufacturer CityKORTRIJK, 8500
Manufacturer CountryBE
Manufacturer Postal Code8500
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameMDRC-1119 WITH PROTECTIVE COVER KIT
Generic NameCLINICAL REVIEW DISPLAY
Product CodePGY
Date Received2017-02-27
Model NumberMDRC-1119 PROTECTIVE COVER KIT
Catalog NumberC9826109
Lot NumberN/A
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBARCO NV
Manufacturer AddressBENELUXPARK 21 KORTRIJK, 8500 BE 8500


Patients

Patient NumberTreatmentOutcomeDate
101. Congenital Not Applicablenomaly 2017-02-27