LEICA M300 10446587

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2017-02-27 for LEICA M300 10446587 manufactured by Leica Microsystems (schweiz) Ag.

Event Text Entries

[68372020] This is a final report. The root cause for the malfunction is maintenance deficiency. The malfunction resulted from an improper service activity by a third party field service engineer (fse), who did not follow the manufacturer's instructions. The service activity occurred half a year prior to the incident. After dismantling the optics carrier fixation to the swing arm of the surgical microscope stand, the third party fse improperly re-used the screws when attaching the optics carrier after service. According to the manufacturer's instructions, the old screws must be exchanged with new screws after service and the manufacturer instructed to never to use screws twice. Re-using the screws twice allowed all three screws to get loose over time, which subsequently caused the optics carrier to fall down unexpectedly.
Patient Sequence No: 1, Text Type: N, H10


[68372021] Leica microsystems ((b)(4)) (b)(4) received a complaint on (b)(6) 2017 from (b)(6) stating that on (b)(6) 2017, the optics carrier of a leica m300 fell onto patient's head during an ear examination.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003974370-2017-00002
MDR Report Key6360227
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2017-02-27
Date of Report2017-02-03
Date of Event2017-02-02
Device Manufacturer Date2007-08-30
Date Added to Maude2017-02-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. ROLAND JEHLE
Manufacturer StreetMAX SCHMIDHEINY STRASSE 201
Manufacturer CityHEERBRUGG, ST. GALLEN 9435
Manufacturer CountrySZ
Manufacturer Postal9435
Manufacturer Phone717263216
Manufacturer G1LEICA INSTRUMENTS (SINGAPORE) PTE LTD
Manufacturer Street12 TEBAN GARDENS CRESCENT
Manufacturer CitySINGAPORE, 608924
Manufacturer CountrySN
Manufacturer Postal Code608924
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLEICA M300
Generic NameSURGICAL MICROSCOPE
Product CodeEPT
Date Received2017-02-27
Model NumberM300
Catalog Number10446587
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerLEICA MICROSYSTEMS (SCHWEIZ) AG
Manufacturer AddressMAX SCHMIDHEINY STRASSE 201 HEERBRUGG, ST. GALLEN 9435 SZ 9435


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-02-27