ABL 700 SERIES ABL 735 *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2005-09-15 for ABL 700 SERIES ABL 735 * manufactured by Radiometer Medical Aps.

Event Text Entries

[411718] The customer performed a decontamination procedure on the analyzer, disregarding the screen warning: "warning: glucose, lactate and chloride electrodes should be replaced with dummy electrodes before proceeding. " the customer left the active electrodes in place during the decontamination procedure damaging the electrodes in question thus causing faulty pt values when the analyzer was subsequently used. Originally the hosp alleged that the faulty values resulted in incorrect treatment contributing to a pt death. The hosp later stated that the reported connection between the results from the analyzer and the death of a pt was a misunderstanding.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002807968-2004-00012
MDR Report Key636044
Report Source05,06
Date Received2005-09-15
Date of Report2004-12-10
Date Mfgr Received2003-02-21
Date Added to Maude2005-09-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactLEIF GUDNITZ
Manufacturer StreetAKANDEVEJ 21,
Manufacturer CityBRONSHOJ
Manufacturer CountryDA
Manufacturer Phone8273228
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Remedial ActionRC
Previous Use Code3
Removal Correction Number3002807-12/10/2004-001C
Event Type3
Type of Report3

Device Details

Brand NameABL 700 SERIES
Generic NameBLOOD GAS ANALYZER WITH ADDITIONAL PARAMETER
Product CodeCCE
Date Received2005-09-15
Model NumberABL 735
Catalog Number*
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key625586
ManufacturerRADIOMETER MEDICAL APS
Manufacturer Address* * *
Baseline Brand NameABL 700 SERIES
Baseline Generic NameBLOOD GAS ANALYZER WITH ADDITIONAL PARAMETER
Baseline Model NoABL 735
Baseline Catalog No*
Baseline ID*


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2005-09-15

© 2026 FDA.report