RANEY SCALP CLIPS 20-1037

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2017-02-27 for RANEY SCALP CLIPS 20-1037 manufactured by Codman & Shurtleff, Inc..

Event Text Entries

[68426591] Udi: (b)(4). Upon completion of the investigation a follow up report will be filed.
Patient Sequence No: 1, Text Type: N, H10


[68426592] As reported by the ous affiliate, during cranial surgery, when a raney scalp clip 10 pack was opened there were 11 in the pack. An alternate pack was used; no delays or adverse events recorded.
Patient Sequence No: 1, Text Type: D, B5


[73355716] The device has been returned. Upon completion of the investigation, a follow-up report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[73693199] Two inner packaging pouches and eleven loose clips were returned for evaluation. While we are unable to definitively confirm that the eleven clips came from a single package, we are confirming the complaint based on the information reported. A review of manufacturing records found no issues related to piece counts. There are several piece count checks in place for this product throughout the processes-the auto bagger deflects rejects according to piece counts and a second scale is to be used to verify 10 counts. Trends will be monitored for this or similar complaints. At present, we consider this complaint to be closed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1226348-2017-10126
MDR Report Key6361518
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2017-02-27
Date of Event2013-11-21
Date Mfgr Received2017-03-21
Date Added to Maude2017-02-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR JAMES KENNEY
Manufacturer Street325 PARAMOUNT DRIVE
Manufacturer CityRAYNHAM MA 02767
Manufacturer CountryUS
Manufacturer Postal02767
Manufacturer Phone5088282726
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRANEY SCALP CLIPS
Generic NameCLIPS, SCALP
Product CodeHBO
Date Received2017-02-27
Returned To Mfg2017-02-28
Catalog Number20-1037
Lot NumberEJ601
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCODMAN & SHURTLEFF, INC.
Manufacturer Address325 PARAMOUNT DRIVE RAYNHAM MA 02767 US 02767


Patients

Patient NumberTreatmentOutcomeDate
10 2017-02-27

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.