COAGULATING BIPOLAR ELECTRODE 28762KB

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-02-28 for COAGULATING BIPOLAR ELECTRODE 28762KB manufactured by Karl Storz Gmbh & Co. Kg Tuttlingen, Germany.

Event Text Entries

[68969162] Evaluation confirmed that the distal end teflon is burned, melted and cracked. Also the shaft is bent, and the proximal end label is illegible. The instrument has a date code of sx(05/2012) and it has been in use for approximately 5 years. Damage is consistent with wear of long use/stress overload.
Patient Sequence No: 1, Text Type: N, H10


[68969163] Allegedly, during a ventriculostomy procedure the doctor noted that the instrument sparked during the procedure. The instrument was immediately replaced and the procedure was completed with no delay. The hospital reported there was no serious injury to patient.
Patient Sequence No: 1, Text Type: D, B5


[71001339] This is a supplemental. I listed the wrong part number in the mdr. I listed 28162kb; it should be 28762kb. I also added udi #. I will make correction.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610617-2017-00019
MDR Report Key6367127
Date Received2017-02-28
Date of Report2017-01-30
Date of Event2017-01-30
Date Mfgr Received2017-01-30
Device Manufacturer Date2012-05-01
Date Added to Maude2017-02-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. SUSIE CHEN
Manufacturer Street2151 E. GRAND AVENUE
Manufacturer CityEL SEGUNDO CA 902455017
Manufacturer CountryUS
Manufacturer Postal902455017
Manufacturer Phone4242188201
Manufacturer G1KARL STORZ GMBH & CO. KG TUTTLINGEN, GERMANY
Manufacturer StreetMITTLESTRASSE 8, 78503
Manufacturer CityTUTTLINGEN, GERMANY,
Manufacturer CountryGM
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCOAGULATING BIPOLAR ELECTRODE
Generic NameCOAGULATING BIPOLAR ELECTRODE
Product CodeGWG
Date Received2017-02-28
Returned To Mfg2017-02-08
Model Number28762KB
Catalog Number28762KB
Lot NumberSX
OperatorPHYSICIAN
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerKARL STORZ GMBH & CO. KG TUTTLINGEN, GERMANY
Manufacturer AddressMITTLESTRASSE 8, 78503 TUTTLINGEN, GERMANY, GM


Patients

Patient NumberTreatmentOutcomeDate
10 2017-02-28

© 2026 FDA.report