[68971147] (b)(4). The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment. Reliability engineering evaluated the device and the reported condition was confirmed. The assignable root cause was determined to be due to improper handling, which is user error/misuse/abuse. If additional information should become available, a supplemental medwatch report will be submitted accordingly.
Patient Sequence No: 1, Text Type: N, H10


[68971148] This is report 4 of 4 for the same event. It was reported from (b)(6) that during an unspecified surgical procedure,it was observed that four air hose devices were leaking. It was not reported if there was a delay in the procedure due to the event. It was not reported if there was a spare device available to complete the procedure. There was patient involvement reported. There were no patient or user injuries reported. It was reported there was no medical intervention or prolonged hospitalization. All available information has been disclosed. If additional information should become available, a supplemental medwatch report will be submitted accordingly.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003862213-2017-10009
MDR Report Key6368080
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2017-03-01
Date of Report2016-10-14
Date of Event2016-10-13
Date Mfgr Received2016-10-14
Device Manufacturer Date2017-09-23
Date Added to Maude2017-03-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMICHAEL COTE
Manufacturer Street1302 WRIGHTS LANE EAST
Manufacturer CityWEST CHESTER PA 19380
Manufacturer CountryUS
Manufacturer Postal19380
Manufacturer Phone6107195000
Manufacturer G1UMKIRCH
Manufacturer StreetIM KIRCHENHURSTLE 4
Manufacturer CityUMKIRCH B. FREIBURG D-79224
Manufacturer CountryGM
Manufacturer Postal CodeD-79224
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDOUBLEAIRHOSE L5M F/SYST SYNTHES
Generic NameINSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Product CodeHSZ
Date Received2017-03-01
Returned To Mfg2017-02-13
Catalog Number519.530
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerUMKIRCH
Manufacturer AddressIM KIRCHENHURSTLE 4 UMKIRCH B. FREIBURG D-79224 GM D-79224


Patients

Patient NumberTreatmentOutcomeDate
10 2017-03-01