MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2017-03-01 for DOUBLEAIRHOSE L5M F/SYST SYNTHES 519.530 manufactured by Umkirch.
[68971147]
(b)(4). The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment. Reliability engineering evaluated the device and the reported condition was confirmed. The assignable root cause was determined to be due to improper handling, which is user error/misuse/abuse. If additional information should become available, a supplemental medwatch report will be submitted accordingly.
Patient Sequence No: 1, Text Type: N, H10
[68971148]
This is report 4 of 4 for the same event. It was reported from (b)(6) that during an unspecified surgical procedure,it was observed that four air hose devices were leaking. It was not reported if there was a delay in the procedure due to the event. It was not reported if there was a spare device available to complete the procedure. There was patient involvement reported. There were no patient or user injuries reported. It was reported there was no medical intervention or prolonged hospitalization. All available information has been disclosed. If additional information should become available, a supplemental medwatch report will be submitted accordingly.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3003862213-2017-10009 |
MDR Report Key | 6368080 |
Report Source | COMPANY REPRESENTATIVE,FOREIG |
Date Received | 2017-03-01 |
Date of Report | 2016-10-14 |
Date of Event | 2016-10-13 |
Date Mfgr Received | 2016-10-14 |
Device Manufacturer Date | 2017-09-23 |
Date Added to Maude | 2017-03-01 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MICHAEL COTE |
Manufacturer Street | 1302 WRIGHTS LANE EAST |
Manufacturer City | WEST CHESTER PA 19380 |
Manufacturer Country | US |
Manufacturer Postal | 19380 |
Manufacturer Phone | 6107195000 |
Manufacturer G1 | UMKIRCH |
Manufacturer Street | IM KIRCHENHURSTLE 4 |
Manufacturer City | UMKIRCH B. FREIBURG D-79224 |
Manufacturer Country | GM |
Manufacturer Postal Code | D-79224 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | DOUBLEAIRHOSE L5M F/SYST SYNTHES |
Generic Name | INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT |
Product Code | HSZ |
Date Received | 2017-03-01 |
Returned To Mfg | 2017-02-13 |
Catalog Number | 519.530 |
Lot Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | UMKIRCH |
Manufacturer Address | IM KIRCHENHURSTLE 4 UMKIRCH B. FREIBURG D-79224 GM D-79224 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-03-01 |