CHATTANOOGA 2738

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2017-03-01 for CHATTANOOGA 2738 manufactured by Djo, Llc.

Event Text Entries

[68635708] Not returned.
Patient Sequence No: 1, Text Type: N, H10


[68635709] Complaint received that alleges "patient,"was on the machine for approx. 15 minutes without incident and there was a sudden surge that shot the electrode off of his leg. When she looked at his leg he had what appears to be 3rd degree electrical burns underneath". Questionnaire not received from customer or clinician. Device not returned to manufacturer for evaluation.
Patient Sequence No: 1, Text Type: D, B5


[74117500] Product was returned for review. The unit and applicator passed functional testing. The electrodes that were sent in expired 10/2016. Two of the electrodes tested above the tolerance level. The high impedance on the electrodes may have contributed to the event.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9616086-2017-00006
MDR Report Key6368197
Report SourceUSER FACILITY
Date Received2017-03-01
Date of Report2017-03-01
Date of Event2017-02-27
Date Mfgr Received2017-04-28
Date Added to Maude2017-03-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag0
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactWILLIAM FISHER
Manufacturer Street1430 DECISION STREET
Manufacturer CityVISTA CA 920819663
Manufacturer CountryUS
Manufacturer Postal920819663
Manufacturer Phone7607313126
Manufacturer G1DJ ORTHOPEDICS DE MEXICO, S.A. DE C.V.
Manufacturer StreetCARRETERA LIBRE TIJUANA TECATE 20230 SUBMETROPOLI EL FLORIDO
Manufacturer CityTIJUANA, 22244
Manufacturer CountryMX
Manufacturer Postal Code22244
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHATTANOOGA
Generic NameINTELECT TRANSPORT COMBO PKG US STD
Product CodeGZI
Date Received2017-03-01
Returned To Mfg2017-04-28
Model Number2738
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerDJO, LLC
Manufacturer Address1430 DECISION STREET VISTA CA 920819663 US 920819663


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-03-01

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.