MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2017-03-01 for CHATTANOOGA 2738 manufactured by Djo, Llc.
[68635708]
Not returned.
Patient Sequence No: 1, Text Type: N, H10
[68635709]
Complaint received that alleges "patient,"was on the machine for approx. 15 minutes without incident and there was a sudden surge that shot the electrode off of his leg. When she looked at his leg he had what appears to be 3rd degree electrical burns underneath". Questionnaire not received from customer or clinician. Device not returned to manufacturer for evaluation.
Patient Sequence No: 1, Text Type: D, B5
[74117500]
Product was returned for review. The unit and applicator passed functional testing. The electrodes that were sent in expired 10/2016. Two of the electrodes tested above the tolerance level. The high impedance on the electrodes may have contributed to the event.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 9616086-2017-00006 |
MDR Report Key | 6368197 |
Report Source | USER FACILITY |
Date Received | 2017-03-01 |
Date of Report | 2017-03-01 |
Date of Event | 2017-02-27 |
Date Mfgr Received | 2017-04-28 |
Date Added to Maude | 2017-03-01 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 0 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | WILLIAM FISHER |
Manufacturer Street | 1430 DECISION STREET |
Manufacturer City | VISTA CA 920819663 |
Manufacturer Country | US |
Manufacturer Postal | 920819663 |
Manufacturer Phone | 7607313126 |
Manufacturer G1 | DJ ORTHOPEDICS DE MEXICO, S.A. DE C.V. |
Manufacturer Street | CARRETERA LIBRE TIJUANA TECATE 20230 SUBMETROPOLI EL FLORIDO |
Manufacturer City | TIJUANA, 22244 |
Manufacturer Country | MX |
Manufacturer Postal Code | 22244 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | CHATTANOOGA |
Generic Name | INTELECT TRANSPORT COMBO PKG US STD |
Product Code | GZI |
Date Received | 2017-03-01 |
Returned To Mfg | 2017-04-28 |
Model Number | 2738 |
Device Availability | R |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | DJO, LLC |
Manufacturer Address | 1430 DECISION STREET VISTA CA 920819663 US 920819663 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2017-03-01 |