TRIGEN 71631161

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2005-09-26 for TRIGEN 71631161 manufactured by Smith & Nephew, Inc. Orthopaedic Div..

Event Text Entries

[386887] It was reported that surgical time was extended when the screws did not align to the holes in the nail properly. The hospital staff attributed the problem to the hexdriver.
Patient Sequence No: 1, Text Type: D, B5


[7848603] Three screws used with the hexdriver were returned. Each of these screws had been cross threaded in the field. It is possible that the cross threading of the screws is responsible for the difficulties experienced.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1020279-2005-00347
MDR Report Key637301
Report Source07
Date Received2005-09-26
Date of Report2005-09-26
Date of Event2005-09-14
Date Mfgr Received2005-09-19
Device Manufacturer Date2004-08-01
Date Added to Maude2005-09-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. ANDY HARDISON,
Manufacturer Street1450 BROOKS ROAD
Manufacturer CityMEMPHIS TN 38116
Manufacturer CountryUS
Manufacturer Postal38116
Manufacturer Phone9013995922
Manufacturer G1SMITH & NEPHEW INC. ORTHOPAEDICS DIV.
Manufacturer Street1450 BROOKS ROAD
Manufacturer CityMEMPHIS TN 38116
Manufacturer CountryUS
Manufacturer Postal Code38116
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRIGEN
Generic NameHEXDRIVER
Product CodeLXM
Date Received2005-09-26
Model NumberNA
Catalog Number71631161
Lot Number04HM05246
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeNA
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key626846
ManufacturerSMITH & NEPHEW, INC. ORTHOPAEDIC DIV.
Manufacturer Address1450 BROOKS RD. MEMPHIS TN 38116 US
Baseline Brand NameTRIGEN
Baseline Generic NameHEXDRIVER
Baseline Model NoNA
Baseline Catalog No71631161
Baseline IDNA


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2005-09-26

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.