HEA 1.2 BEADCHIP KIT, SLIDE 800-20202-08

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-03-06 for HEA 1.2 BEADCHIP KIT, SLIDE 800-20202-08 manufactured by Bioarray Solutions Ltd..

Event Text Entries

[69411616] The molecular presence of both guanine and adenine at c. 125 in the absence of the silencing polymorphism at ackr1 -67 and weakening polymorphism at c. 265 would normally represent a fy*01/02 (fya+b+) individual 1,2. However, a novel polymorphism in the gata motif at ackr1 -69t>c, two basepairs upstream of the canonical -67t>c gata silencing mutation (fy*02n. 01)3, would most likely result in the silencing of the linked duffy allele. This novel mutation may explain the serological fya- result while molecularly the result is fya+ as in beadchips heac1556_5 and head0316_2. A second novel polymorphism in the 5'utr was also found at ackr1 -248t>a. Silencing mutations are listed as limitations in the precise type hea package insert.
Patient Sequence No: 1, Text Type: N, H10


[69411617] The customer reported a possible discrepancy. The donor is fya+ using the bioarray hea molecular beadchip kit; serology results were fya-.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005967741-2017-00015
MDR Report Key6380201
Date Received2017-03-06
Date of Report2017-03-06
Date of Event2017-02-09
Date Mfgr Received2017-02-09
Device Manufacturer Date2016-06-09
Date Added to Maude2017-03-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. LUZ VILLAMIZAR
Manufacturer Street35 TECHNOLOGY DRIVE SUITE 100
Manufacturer CityWARREN NJ 07059
Manufacturer CountryUS
Manufacturer Postal07059
Manufacturer Phone9084449591
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameHEA 1.2 BEADCHIP KIT, SLIDE
Generic NameHEA 1.2 BEADCHIP KIT, SLIDE
Product CodePEP
Date Received2017-03-06
Model Number800-20202-08
Catalog Number800-20202-08
Lot Number16-313
Device Expiration Date2017-03-31
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBIOARRAY SOLUTIONS LTD.
Manufacturer Address35 TECHNOLOGY DRIVE SUITE 100 WARREN NJ 07059 US 07059


Patients

Patient NumberTreatmentOutcomeDate
10 2017-03-06