STAMMBERGER SINUS DRESSING RR 650

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-03-06 for STAMMBERGER SINUS DRESSING RR 650 manufactured by Arthrocare Corporation.

Event Text Entries

[69498527]
Patient Sequence No: 1, Text Type: N, H10


[69498528] It was reported that during a case it was noticed that the stammberger sinus dressing did not have a syringe in the packaging, the case was completed by using another vendors product. No further complications were reported.
Patient Sequence No: 1, Text Type: D, B5


[74192397] The product, intended for use in treatment, was returned for evaluation. There was a relationship between the product and the reported event. Visual observation of the returned sealed inner tray confirmed the molded cannula was missing. A functional evaluation was not required as the complaint was not related to the functionality of the product. The complaint regarding the missing? Syringe? Has been verified. The root cause of this event is due to a workmanship related issue.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3006524618-2017-00055
MDR Report Key6381106
Date Received2017-03-06
Date of Report2017-04-24
Date of Event2017-02-20
Date Mfgr Received2017-02-24
Date Added to Maude2017-03-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJIM GONZALES
Manufacturer Street7000 WEST WILLIAM CANNON DRIVE
Manufacturer CityAUSTIN TX 78735
Manufacturer CountryUS
Manufacturer Postal78735
Manufacturer Phone5123585706
Manufacturer G1ARTHROCARE CORPORATION
Manufacturer Street7000 W. WILLIAM CANNON
Manufacturer CityAUSTIN TX 78735
Manufacturer CountryUS
Manufacturer Postal Code78735
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTAMMBERGER SINUS DRESSING
Generic NameSPLINT, INTRANASAL SEPTAL
Product CodeLYA
Date Received2017-03-06
Returned To Mfg2017-03-06
Catalog NumberRR 650
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerARTHROCARE CORPORATION
Manufacturer Address7000 W. WILLIAM CANNON AUSTIN TX 78735 US 78735


Patients

Patient NumberTreatmentOutcomeDate
10 2017-03-06

© 2024 FDA.report