DIMENSION VISTA 1500

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-03-06 for DIMENSION VISTA 1500 manufactured by Siemens Healthcare Diagnostics Inc..

Event Text Entries

[69212797] The customer contacted a siemens customer care center (ccc) specialist. The ccc was granted permission to remotely connect to the customer's instrument. The ccc reviewed the data provided. The customer's quality control (qc) was in range at the time of the event. The customer cleaned aliquot probe and drain. The customer ran check 1, resulting within range. The instrument was reset. A precision study was performed, resulting within range. The cause of the discordant, falsely depressed urinary cerebrospinal fluid protein result is unknown. The instrument is performing according to specifications. No further evaluation of the device is required.
Patient Sequence No: 1, Text Type: N, H10


[69212798] A discordant, falsely depressed urinary cerebrospinal fluid protein result was obtained on a patient sample on a dimension vista 1500 instrument. The initial result was reported out to the physician(s). The customer repeated the same sample on the same dimension vista instrument, resulting higher. The customer repeated the same sample on an alternate dimension vista instrument, resulting higher than the original result. The customer issued a corrected report to the physician(s). There are no known reports of patient intervention or adverse health consequences due to the discordant, falsely depressed urinary cerebrospinal fluid protein result.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2517506-2017-00231
MDR Report Key6382438
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2017-03-06
Date of Report2017-03-06
Date of Event2017-02-09
Date Mfgr Received2017-02-09
Device Manufacturer Date2013-01-24
Date Added to Maude2017-03-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactTIMOTHY RICE
Manufacturer Street511 BENEDICT AVE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242406
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC
Manufacturer StreetREGISTRATION NUMBER: 1226181 101 SILVERMINE ROAD
Manufacturer CityBROOKFIELD CT 06804
Manufacturer CountryUS
Manufacturer Postal Code06804
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameDIMENSION VISTA 1500
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeJGQ
Date Received2017-03-06
Model NumberDIMENSION VISTA 1500
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No0
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Address500 GBC DRIVE PO BOX 6101 NEWARK DE 197146101 US 197146101

Device Sequence Number: 1

Brand NameDIMENSION VISTA 1500
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeJJE
Date Received2017-03-06
Model NumberDIMENSION VISTA 1500
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Address500 GBC DRIVE PO BOX 6101 NEWARK DE 197146101 US 197146101


Patients

Patient NumberTreatmentOutcomeDate
10 2017-03-06

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.