VALO GRAND 5972

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2017-03-06 for VALO GRAND 5972 manufactured by Ultradent Products, Inc..

Event Text Entries

[69202484] Filing out of an abundance of caution due to the fact the valo was used during restorations which later turned out to be the restoration material used and cure times stated on the manufactures restoration material. Dr. Did not want to return light as he indicated the light functioned as intended. No device malfunction reported.
Patient Sequence No: 1, Text Type: N, H10


[69202485] Doctor has been using activa for about 2 years now, with no issues. Recently had 4 patients develop abscesses associated with small mo/do restorations placed approximately 6 months ago. The only change in protocol within the last six months is the recently purchased curing light. Doctor went from a valo corded to a valo cordless grand. Used the cordless on the fastest (curing) setting due to the fact that speed is key for pediatric patients. Email came in asking if there were any output differences between the corded and the cordless.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1718912-2017-00002
MDR Report Key6382979
Report SourceHEALTH PROFESSIONAL
Date Received2017-03-06
Date of Report2017-03-06
Date of Event2017-02-02
Date Mfgr Received2017-02-02
Device Manufacturer Date2016-06-26
Date Added to Maude2017-03-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. RONA MURPHY
Manufacturer Street505 WEST 10200 SOUTH
Manufacturer CitySOUTH JORDAN UT 84095
Manufacturer CountryUS
Manufacturer Postal84095
Manufacturer Phone8015534200
Manufacturer G1ULTRADENT PRODUCTS, INC.
Manufacturer Street505 WEST 10200 SOUTH
Manufacturer CitySOUTH JORDAN UT 84095
Manufacturer CountryUS
Manufacturer Postal Code84095
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVALO GRAND
Generic NameDENTAL CURING LIGHT
Product CodeEBZ
Date Received2017-03-06
Catalog Number5972
Lot NumberBD44B
OperatorDENTIST
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerULTRADENT PRODUCTS, INC.
Manufacturer Address505 WEST 10200 SOUTH SOUTH JORDAN UT 84095 US 84095


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-03-06

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.