HYPERBARIC CHAMBER 3600ER

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a other,user facility report with the FDA on 2017-03-09 for HYPERBARIC CHAMBER 3600ER manufactured by Sechrist Industries, Inc..

Event Text Entries

[69564737] Sechrist technician was dispatched onsite on 2/26/2016. Technician performed max pressure test, checked/inspected control panel tubings and other components for possible leaks. Performed chamber control system calibration iaw section 11. 5 (electronic control module manual). Performed & completed operational/verification. (b)(4) staff and technician ran chamber for 90 minutes w/rs of 1. 5 & ps of 2. 4 ata, chamber maintained pressure for duration of test. A review of the device history record (dhr) shows no indication that there were any relevant discrepancies during manufacturing and found no nonconformance that could cause or contribute to the reported issue.
Patient Sequence No: 1, Text Type: N, H10


[69564738] It was reported that hyperbaric chamber pressure drops from 2. 4ata to 2. 3ata. No patient injury was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2020676-2017-00003
MDR Report Key6393331
Report SourceOTHER,USER FACILITY
Date Received2017-03-09
Date of Report2016-02-23
Date of Event2016-02-23
Date Mfgr Received2016-02-23
Device Manufacturer Date2010-02-01
Date Added to Maude2017-03-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. VICTOR ARELLANO
Manufacturer Street4225 E. LA PALMA AVENUE
Manufacturer CityANAHEIM CA 92807
Manufacturer CountryUS
Manufacturer Postal92807
Manufacturer Phone7145798400
Manufacturer G1SECHRIST INDUSTRIES, INC.
Manufacturer Street4225 E. LA PALMA AVENUE
Manufacturer CityANAHEIM CA 92807
Manufacturer CountryUS
Manufacturer Postal Code92807
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHYPERBARIC CHAMBER
Generic NameCHAMBER, HYPERBARIC
Product CodeCBF
Date Received2017-03-09
Model Number3600ER
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSECHRIST INDUSTRIES, INC.
Manufacturer Address4225 E. LA PALMA AVENUE ANAHEIM CA 92807 US 92807


Patients

Patient NumberTreatmentOutcomeDate
10 2017-03-09

© 2026 FDA.report