MAYO-HEGAR NEEDLE HOLDER STANDARD TC N/A 51-6511

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2017-03-10 for MAYO-HEGAR NEEDLE HOLDER STANDARD TC N/A 51-6511 manufactured by Biomet Microfixation.

Event Text Entries

[69651648] Ak is the date code etched on the product which identifies the item as being manufactured between january and june of 1995, however the lot number is unable to determined. A visual inspection revealed moderate signs of wear including minor scratches and spots of discoloration. A small piece of the knurled jaw was broken off. The most-likely cause for the complaint was determined to be excessive force. There are no indications of manufacturing defects. This report is late as it was part of a retrospective review based on enhancements made to our reporting policies.
Patient Sequence No: 1, Text Type: N, H10


[69651649] It was originally reported this instrument "doesn't latch anymore. " the event occurred during surgery with no delay or harm to the patient. However the product evaluation identified the product was broken at the tip. See h10 narrative for the product evaluation details.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0001032347-2017-00145
MDR Report Key6396979
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2017-03-10
Date of Report2016-12-09
Date Mfgr Received2016-12-09
Date Added to Maude2017-03-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. MICHELLE COLE
Manufacturer Street1520 TRADEPORT DRIVE
Manufacturer CityJACKSONVILLE FL 32218
Manufacturer CountryUS
Manufacturer Postal32218
Manufacturer Phone9047414400
Manufacturer G1BIOMET MICROFIXATION
Manufacturer Street1520 TRADEPORT DRIVE
Manufacturer CityJACKSONVILLE FL 32218
Manufacturer CountryUS
Manufacturer Postal Code32218
Single Use3
Previous Use Code3
Removal Correction NumberN/A
Event Type3
Type of Report3

Device Details

Brand NameMAYO-HEGAR NEEDLE HOLDER STANDARD TC
Generic NameNEEDLE HOLDER
Product CodeHXK
Date Received2017-03-10
Returned To Mfg2016-10-28
Model NumberN/A
Catalog Number51-6511
Lot NumberUNKNOWN
ID NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBIOMET MICROFIXATION
Manufacturer Address1520 TRADEPORT DRIVE JACKSONVILLE FL 32218 US 32218


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2017-03-10