PROBEAT-V

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-03-11 for PROBEAT-V manufactured by Hitachi, Ltd., Healthcare Hitachi Works.

Event Text Entries

[70017827] A parameter in the positioning image analysis system (pias) for isocenter position #1 in treatment room 2 was mistakenly changed during the software modification work on (b)(6) 2016. A service and maintenance engineer discovered the mis-setting of the parameter on (b)(6) 2017, and it was corrected immediately. According to the treatment record after (b)(6) 2016, one patient was found to receive treatments around the shoulder at isocenter position #1. The fourteen fractions out of twenty-five were applied to the patient under the faulty condition starting from (b)(6) 2016. The incorrect parameter could have affected the patient positioning accuracy. As a result of investigation, the amount of deviation is estimated in the range of 0. 24cm to 1. 36cm, varying by each fraction. No adverse event for the patient has been reported. This issue has been resolved and the system is currently working correctly.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003993895-2017-00001
MDR Report Key6398200
Date Received2017-03-11
Date of Report2017-03-11
Date of Event2017-02-10
Date Mfgr Received2017-02-10
Device Manufacturer Date2016-01-27
Date Added to Maude2017-03-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. MIKIMOTO TAKAMATSU
Manufacturer Street50 PROSPECT AVE
Manufacturer CityNORTH TARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145246636
Manufacturer G1HITACHI, LTD., HEALTHCARE HITACHI WORKS
Manufacturer Street3-1-1 SAIWAI-CHO
Manufacturer CityHITACHI-SHI, IBARAKI 317-8511
Manufacturer CountryJA
Manufacturer Postal Code317-8511
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NamePROBEAT-V
Generic NamePROTON BEAM THERAPY SYSTEM
Product CodeLHN
Date Received2017-03-11
Model NumberPROBEAT-V
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHITACHI, LTD., HEALTHCARE HITACHI WORKS
Manufacturer Address3-1-1 SAIWAI-CHO HITACHI-SHI, IBARAKI 317-8511 JA 317-8511


Patients

Patient NumberTreatmentOutcomeDate
10 2017-03-11