BLOODGAS ANALYZER ABL735 *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2005-10-10 for BLOODGAS ANALYZER ABL735 * manufactured by Radiometer Medical Aps.

Event Text Entries

[414594] Radiometer received the following complaint from our german distributor on 09/2005 at 8:38 in the morning: we received information from our technician about problems on measurement of the parameter ph on the analyzer abl725. He reported that her doctor called from the hospital called him and described the following situation: last night they had a faulty measurement on parameter ph. The blood sample was taken from a neonate and the parameter was clearly too low. The baby is female born in 09/2005 (week gestation 37) birth weights 3700g, with the admission diagnosis spesis with pulmonary-involvement. The new born was buffered with 12mmol sodium bicarbonate. After that, one of the doctors took a blood sample from himself and his measure value for ph was so low that he normally should be admitted to intensive care unit. Half an hour later, the measured values were back to normal. It was reported that the problem with too low measured ph values had appeared several times the last months. Unfortunately, the customer did not measure qc's in this regard. The analyzer did not show any error messages while measuring too deep on ph and also no error messages on calibrations. Our technician replaced the ph electrode (e777; lot pl-02) with a new one. The ref-electrode (e1001;lot 49-5) is still placed in the analyzer. The concerned electrode is on theway back to us. On september doctor was asked about the condition of the baby and he answered that the neonate has not been harmed and the current status of the baby is absolutely fine.
Patient Sequence No: 1, Text Type: D, B5


[433029] A blood sample was taken from a patient and the reported ph value was very low. The doctor treated the patient according to the measured ph-values without therapeutic success. After that, a new sample from a healthy adult was measured; this gave the same low ph value. Half an hour later, the reported values were back to normal. The site has reported that the problem with too low ph values had appeared several times over the last months. The customer did not use qc's in this case. The analyzer did not show any error messages while measuring too low on ph or any error messages on calibrations. No adverse outcome, the patient is reported in fine health.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002807968-2005-00010
MDR Report Key640527
Report Source01,05,06
Date Received2005-10-10
Date of Report2005-10-06
Date of Event2005-09-06
Date Mfgr Received2005-09-06
Device Manufacturer Date1999-06-01
Date Added to Maude2005-10-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactLEIF GUDNITZ
Manufacturer StreetAKANDEVEJ 21
Manufacturer CityBRONSHOJ
Manufacturer CountryDA
Manufacturer Phone8273228
Manufacturer G1RADIOMETER MEDICAL APS
Manufacturer StreetAKANDEVEJ 21
Manufacturer CityBRONSHOJ
Manufacturer CountryDA
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBLOODGAS ANALYZER
Generic NameABL735
Product CodeCCE
Date Received2005-10-10
Model NumberABL735
Catalog Number*
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device Eval'ed by MfgrR
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key630036
ManufacturerRADIOMETER MEDICAL APS
Manufacturer Address* * *
Baseline Brand NameABL735
Baseline Generic NameBLOOD GAS ANALYSER WITH ADDITIONAL PARAMETERS
Baseline Model NoABL735
Baseline Catalog NoNA
Baseline IDNA


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2005-10-10