CWP 100 RO, CWP, 28 LPM, 104H

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-03-15 for CWP 100 RO, CWP, 28 LPM, 104H manufactured by Mar Cor Purification, Inc..

Event Text Entries

[69978621]
Patient Sequence No: 1, Text Type: N, H10


[69978622] For about 2 weeks, it was reported that there had been repetitive low level inlet pressure alarms overnight and the green operations indicator light would not come on. Due to these issues, the water purification unit could not be used and the back-up deionization system was used to purify the water. The system was serviced by the manufacturer 4 times during the 2 week period. Services performed by the manufacturer included changing the pre-treatment back wash timer settings, replacing the low level switch for the hot water tank, tightening the jumper bar, and replacing valve 64. Reportedly, the water purification unit has not had any issues since the last service when the jumper bar was tightened. Reportedly, no harm occurred to patients during the 2 week period.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number6405704
MDR Report Key6405704
Date Received2017-03-15
Date of Report2017-02-22
Date of Event2017-01-13
Report Date2017-02-22
Date Reported to FDA2017-02-22
Date Reported to Mfgr2017-02-22
Date Added to Maude2017-03-15
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCWP 100
Generic NameSUBSYSTEM, WATER PURIFICATION
Product CodeFIP
Date Received2017-03-15
Model NumberRO, CWP, 28 LPM, 104H
ID Number100474
Device Availability*
Device Age16 MO
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMAR COR PURIFICATION, INC.
Manufacturer Address14550 28TH AVENUE NORTH MINNEAPOLIS MN 55447 US 55447


Patients

Patient NumberTreatmentOutcomeDate
10 2017-03-15

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.