IVC BARIATRIC BED 9153641598 NA:BAR600IVC BAR600IVC

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2017-03-15 for IVC BARIATRIC BED 9153641598 NA:BAR600IVC BAR600IVC manufactured by Invacare Taylor St..

Event Text Entries

[69970819] The motors, junction box and pendant were returned. The bed frame was not returned. An evaluation of the components has not been completed at this time, therefore the cause of the fire could not be determined. The dealer stated the bed was set up in the facility and has been in operation until the alleged event. The bed is 11 years old. The dealer has not ordered a replacement. Should additional information become available, a supplemental record will be filed.
Patient Sequence No: 1, Text Type: N, H10


[69970820] The tbm called stated the bed had a fire. The tbm stated no property damage no injury. There was a patient in the bed when the event happened but was not hurt.
Patient Sequence No: 1, Text Type: D, B5


[73304202] The following components were returned for an expanded evaluation: 6 function control box, 6 way hand pendant, foot motor, hi/lo motor and head motor. The complaint was confirmed for the failed control box. The reported issued of a fire could not be duplicated as the control box was not functionally tested due to the received condition. There was discoloration and deformation consistent with heat or smoke damage near the hi/lo and bed molex motor connectors. The control box housing is comprised of v0 rated material and will self-extinguish once the heat source is removed. The root cause can be attributed to a capacitor failure. The control box was dated for september 6, 2005. The product is approximately 12 years old which is well beyond its end of life. An event review request has been submitted to address this issue.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1031452-2017-00034
MDR Report Key6405751
Report SourceDISTRIBUTOR
Date Received2017-03-15
Date of Report2017-02-13
Date Mfgr Received2017-04-12
Device Manufacturer Date2005-09-01
Date Added to Maude2017-03-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR JASON FIEST
Manufacturer StreetONE INVACARE WAY
Manufacturer CityELYRIA OH 44036
Manufacturer CountryUS
Manufacturer Postal44036
Manufacturer Phone8003336900
Manufacturer G1INVACARE TAYLOR ST.
Manufacturer Street1200 TAYLOR ST.
Manufacturer CityELYRIA OH 44036
Manufacturer CountryUS
Manufacturer Postal Code44036
Single Use3
Remedial ActionIN
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIVC BARIATRIC BED 9153641598
Generic NameBED, BARIATRIC
Product CodeOSI
Date Received2017-03-15
Returned To Mfg2017-03-14
Model NumberNA:BAR600IVC
Catalog NumberBAR600IVC
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerINVACARE TAYLOR ST.
Manufacturer Address1200 TAYLOR STREET ELYRIA OH 44036 US 44036


Patients

Patient NumberTreatmentOutcomeDate
10 2017-03-15

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.