API 20 C AUX 25 STRIPS+25MEDIA 20210

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,literatur report with the FDA on 2017-03-15 for API 20 C AUX 25 STRIPS+25MEDIA 20210 manufactured by Biomerieux Sa.

Event Text Entries

[70105500] On (b)(6) 2017, an internal complaint was initiated following a scientific publication entitled, "laboratory identification of candida auris," on potential misidentification of c. Auris strains on api 20 c aux 25 strips+25media (ref. 20210). Api 20 c aux is a system for the precise identification of the most frequently encountered yeasts. The complete list of those species that it is possible to identify with this system is given in the identification table at the end of this package insert. The api 20 c aux strip consists of 20 cupules containing dehydrated substrates which enable the performance of 19 assimilation tests. The cupules are inoculated with a semi-solid minimal medium. The yeasts will only grow if they are capable of utilizing each substrate as the sole carbon source. As this internal complaint is related to a scientific publication, to date, there is no evidence of a performance issue and there is no patient outcomes. Based on the information provided there is no evidence of any impact on a patient or adverse even occurring, or whether there was a delay in treatment based on this event. An internal biom? Rieux investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2017-00051
MDR Report Key6407514
Report SourceHEALTH PROFESSIONAL,LITERATUR
Date Received2017-03-15
Date of Report2017-11-30
Date Mfgr Received2017-11-06
Date Added to Maude2017-03-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. ELLEN WELTMER
Manufacturer Street595 ANGLUM ROAD
Manufacturer CitySAINT LOUIS MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer G1BIOM
Manufacturer Street5, RUE DES AQUEDUCS CRAPONNE, FR 69290
Manufacturer CountryFR
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAPI 20 C AUX 25 STRIPS+25MEDIA
Generic NameAPI 20 C AUX 25 STRIPS+25MEDIA
Product CodeJSS
Date Received2017-03-15
Model Number20210
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer AddressCHEMIN DE L ORME MARCY L ETOILE, RHONE 69280 FR


Patients

Patient NumberTreatmentOutcomeDate
10 2017-03-15

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.