SW-180

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,foreign,user faci report with the FDA on 2017-03-17 for SW-180 manufactured by Ito Co., Ltd. Tsukuba Factory.

Event Text Entries

[70249580] The patient had been receiving treatment using the subject device with 40 w output in continuous mode. The subject device burnt the patient on the hip during the thirteenth treatment. According to the hospital, the burn had fully healed by (b)(6) 2014 when ito received the information on the event.
Patient Sequence No: 1, Text Type: D, B5


[72687668] Ito examined the device history record for the device involved in the adverse event. There were no problems observed during the manufacturing or testing noted in the dhr. Only the oscillator unit was returned to ito. Ito conducted the evaluation of the returned oscillator unit but there was no output from the unit. Ito received a report describing misuse by the user and documented it in the investigation report on the subject device (s/n (b)(4)). Ito reminded the user of the correct use of the device, as described in the user manual.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9614750-2017-00007
MDR Report Key6412980
Report SourceDISTRIBUTOR,FOREIGN,USER FACI
Date Received2017-03-17
Date of Report2017-03-17
Date of Event2014-09-20
Date Mfgr Received2014-10-17
Device Manufacturer Date2014-01-30
Date Added to Maude2017-03-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. KENNETH BLOCK
Manufacturer Street1201 RICHARDSON DR. SUITE 160
Manufacturer CityRICHARDSON TX 75080
Manufacturer CountryUS
Manufacturer Postal75080
Manufacturer Phone9724809554
Manufacturer G1ITO CO., LTD. TSUKUBA FACTORY
Manufacturer Street3-39-14 UZURANO AMI-MACHI
Manufacturer CityINASHIKI-GUN, IBARAKI 300-0341
Manufacturer CountryJA
Manufacturer Postal Code300-0341
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSW-180
Generic NameDIATHERMY, SHORTWAVE, FOR USE IN APPLYING THERAPEUTIC DEEP HEAT
Product CodeIMJ
Date Received2017-03-17
Returned To Mfg2014-12-24
Model NumberSW-180
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerITO CO., LTD. TSUKUBA FACTORY
Manufacturer Address3-39-14 UZURANO AMI-MACHI INASHIKI-GUN, IBARAKI 300-0341 JA 300-0341


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-03-17

© 2024 FDA.report