MONOMAX VIOLET 1(4)150CM HRT48 LOOP(M B0041222

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-03-17 for MONOMAX VIOLET 1(4)150CM HRT48 LOOP(M B0041222 manufactured by B.braun Surgical Sa.

Event Text Entries

[70517275] (b)(4). Manufacturing site evaluation: evaluation on-going.
Patient Sequence No: 1, Text Type: N, H10


[70517276] Country of complaint: (b)(6). It is reported that the thread breaks during knotting.
Patient Sequence No: 1, Text Type: D, B5


[73482925] Samples received: 1 unopened pouch. Analysis and results: there are no previous complaints of this batch. Manufactured and distributed (b)(4) units of this batch. There are no units in stock. Received one closed sample. Tightness test to the sample received has been performed and the unit is tight. We have tested the knot pull tensile strength of the sample received and the result fulfils the requirements. Xi= 7. 19 kgf (requirements: 5. 18 kgf in average and 2. 59 kgf in minimum). Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfil b. Braun surgical requirements. Remarks: when working with monomax suture material, great care should be taken to avoid any crushing or crimping damage of the monofilament by instruments such as forceps or needle holders. Actions on product distributed of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product. Corrective/preventive actions: according to internal procedures, there is no need to establish corrective or preventive actions. Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3003639970-2017-00149
MDR Report Key6415195
Date Received2017-03-17
Date of Report2017-03-27
Date Facility Aware2017-03-13
Date Mfgr Received2017-02-16
Date Added to Maude2017-03-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. NICOLE BROYLES
Manufacturer Street615 LAMBERT POINTE DRIVE
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3145515988
Manufacturer G1B.BRAUN SURGICAL SA
Manufacturer Street121 CARRETERA DE TERRASSA
Manufacturer CityRUBI, BARCELONA 08191
Manufacturer CountrySP
Manufacturer Postal Code08191
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMONOMAX VIOLET 1(4)150CM HRT48 LOOP(M
Generic NameSUTURES
Product CodeNWJ
Date Received2017-03-17
Returned To Mfg2017-03-16
Model NumberB0041222
Catalog NumberB0041222
Lot Number116412
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerB.BRAUN SURGICAL SA
Manufacturer Address121 CARRETERA DE TERRASSA RUBI, BARCELONA 08191 SP 08191


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-03-17

© 2026 FDA.report