ARCHITECT AFP 03P36-26

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2017-03-20 for ARCHITECT AFP 03P36-26 manufactured by Abbott Ireland.

Event Text Entries

[70412199] An evaluation is in process. A follow-up report will be submitted when the evaluation is complete.
Patient Sequence No: 1, Text Type: N, H10


[70412200] The customer observed a falsely elevated afp result for one patient while using the architect afp assay. The customer provided the following data (unit of measure not provided): initial result: about 200, retest result: 2-3. There was no adverse impact to patient management reported.
Patient Sequence No: 1, Text Type: D, B5


[75436216] Further investigation of the customer issue included a review of the complaint text, trouble shooting of the analyzer, a search for similar complaints, and a review of labeling. Return material was not available. Review of the analyzer was completed and did not identify any issues. A trending review did not identify an adverse trend. Labeling was reviewed and found to be adequate. Based on the available information a deficiency of the architect afp reagent, list 03p36-26, was not identified.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3008344661-2017-00022
MDR Report Key6417501
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2017-03-20
Date of Report2017-05-15
Date Mfgr Received2017-04-19
Date Added to Maude2017-03-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT IRELAND
Manufacturer StreetDIAGNOSTICS DIVISION FINISKLIN BUSINESS PARK
Manufacturer CitySLIGO NA
Manufacturer CountryEI
Manufacturer Postal CodeNA
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameARCHITECT AFP
Generic NameALPHA-FETOPROTEIN
Product CodeLOK
Date Received2017-03-20
Catalog Number03P36-26
Lot NumberUNKNOWN
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT IRELAND
Manufacturer AddressDIAGNOSTICS DIVISION FINISKLIN BUSINESS PARK SLIGO NA EI NA


Patients

Patient NumberTreatmentOutcomeDate
10 2017-03-20