BUSH URETERAL ILLUMINATING CATHETER SET N/A 084120

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2017-03-21 for BUSH URETERAL ILLUMINATING CATHETER SET N/A 084120 manufactured by Cook Inc.

Event Text Entries

[70519529] (b)(4). The event is currently under investigation. A follow up report will be send upon conclusion.
Patient Sequence No: 1, Text Type: N, H10


[70519530] It was reported by the user facility that during a robotic procedure, the bush ureteral illuminating catheter set was very hot at the entry point where the light cord connects to the lightbox. The nurse indicated she smelled a burning smell, and it was determined that the brass fitting appeared to be overly hot. The power setting was set to 60% on the lightbox. The brand of the light box was luxtec. The customer did not specify whether or not the light source was on the entire time or if it was dimmed when not in use. The customer did state that the overheating occurred towards the end of the procedure and that they turned off the light to the catheter and continued procedure without it. There was no harm to either the patient or the staff. There were no additional procedure necessary because of the brass fitting being? Overly hot?. No further information was provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1820334-2017-00429
MDR Report Key6420269
Report SourceCOMPANY REPRESENTATIVE
Date Received2017-03-21
Date of Report2017-08-14
Date of Event2017-02-15
Date Mfgr Received2017-07-31
Date Added to Maude2017-03-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. RITA HARDEN
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal47404
Manufacturer Phone8123392235
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBUSH URETERAL ILLUMINATING CATHETER SET
Generic NameFCS CATHETER, LIGHT, FIBEROPTIC, GLASS, URETERAL
Product CodeFCS
Date Received2017-03-21
Model NumberN/A
Catalog Number084120
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerCOOK INC
Manufacturer Address750 DANIELS WAY BLOOMINGTON IN 47404 US 47404


Patients

Patient NumberTreatmentOutcomeDate
10 2017-03-21

© 2026 FDA.report