MEDRAD SWABBABLE VALVE TRANSFER SET 60729202 SVTS

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-03-22 for MEDRAD SWABBABLE VALVE TRANSFER SET 60729202 SVTS manufactured by Bayer Medical Care.

Event Text Entries

[70881159] Product analysis received and examined the returned product. Visual examination of the returned product confirmed the presence of particulate inside the tubing. Examination of the particulate found it to be partially embedded within the tubing wall. The particulate measures approximately 1. 25 cm in length. Our investigation determined that the particulate noted in the syringe was introduced during the manufacturing process. The syringe kit instructions for use instructs the user to "visually inspect contents and package before each use". This visual inspection is to ensure particulates are noticed and mitigated, which is what occurred in this reported instance.
Patient Sequence No: 1, Text Type: N, H10


[70881160] The following information was obtained from the customer: the customer reported black foreign body inside the transfer set tubing. No injury or adverse event was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2520313-2017-00018
MDR Report Key6424985
Date Received2017-03-22
Date of Report2017-03-22
Date of Event2017-02-24
Date Mfgr Received2017-02-24
Date Added to Maude2017-03-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. SUSAN SHERWIN
Manufacturer Street1 BAYER DRIVE
Manufacturer CityINDIANOLA PA 15051
Manufacturer CountryUS
Manufacturer Postal15051
Manufacturer Phone7249408678
Manufacturer G1BAYER MEDICAL CARE
Manufacturer Street1 BAYER DRIVE
Manufacturer CityINDIANOLA PA 15051
Manufacturer CountryUS
Manufacturer Postal Code15051
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameMEDRAD SWABBABLE VALVE TRANSFER SET
Generic NameTUBING, FLUID DELIVERY
Product CodeFPK
Date Received2017-03-22
Returned To Mfg2017-03-16
Model Number60729202
Catalog NumberSVTS
Lot NumberUNKNOWN
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBAYER MEDICAL CARE
Manufacturer Address1 BAYER DRIVE INDIANOLA PA 15051 US 15051


Patients

Patient NumberTreatmentOutcomeDate
10 2017-03-22

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.