ETEST? TICARCILLIN CLAVULANIC ACID 522658

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-03-29 for ETEST? TICARCILLIN CLAVULANIC ACID 522658 manufactured by Biom?rieux Sa.

Event Text Entries

[71325087] A customer in the united states reported a discrepant result (out-of-range-low) for eubacterium lentum atcc 43055, associated with etest? Ticarcillin clavulanic acid tic/clav-con2 tlc 256 us f100. The customer reported obtaining an out of range lower mic. The expected values between 16 - 64 and the mic result obtain was 4 - 8. There is no indication or report from the laboratory that the discrepant result led to any adverse event related to any patient's state of health. There may be a potential for adverse event if the event were to reoccur while testing a patient isolate; therefore this event is being reported as a malfunction. Culture submittal was requested by biom? Rieux for internal investigation. An internal biom? Rieux investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2017-00056
MDR Report Key6441962
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2017-03-29
Date of Report2017-08-23
Date Mfgr Received2017-07-28
Date Added to Maude2017-03-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. ELLEN WELTMER
Manufacturer Street595 ANGLUM ROAD
Manufacturer CitySAINT LOUIS MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer G1BIOM
Manufacturer Street5, RUE DES AQUEDUCS CRAPONNE, FR 69290
Manufacturer CountryFR
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameETEST? TICARCILLIN CLAVULANIC ACID
Generic NameETEST? TICARCILLIN CLAVULANIC ACID
Product CodeJWY
Date Received2017-03-29
Catalog Number522658
Lot Number1004598840
Device Expiration Date2019-01-19
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOM?RIEUX SA
Manufacturer AddressCHEMIN DE L ORME MARCY L ETOILE, RHONE 69280 FR


Patients

Patient NumberTreatmentOutcomeDate
10 2017-03-29