LEKSELL GAMMA KNIFE ICON 11.1

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,other report with the FDA on 2017-03-30 for LEKSELL GAMMA KNIFE ICON 11.1 manufactured by Elekta Instrument Ab.

Event Text Entries

[71214377] The matchport is mounted in the gantry module. For version 11. 1, the matchport device must be configured in a special way. This special configuration of the matchport device is described in the installation manual but not in the corrective maintenance manual. After the exchange of the gantry module, the matchport device may thus not be configured properly resulting in risk to the service technician or malfunction of the gantry. If the service technician does a kv generator calibration after the exchange, the technician will find out that the configuration is missing because they would not be able to perform the calibration.
Patient Sequence No: 1, Text Type: N, H10


[71214378] When replacing a faulty gantry module on a new installation, it was discovered that settings controlling the kv generator using lantronix/matchport were lost as they are stored locally in the gantry module.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9612186-2017-00004
MDR Report Key6445391
Report SourceCOMPANY REPRESENTATIVE,OTHER
Date Received2017-03-30
Date of Report2017-03-30
Date Mfgr Received2017-03-02
Date Added to Maude2017-03-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPMS
Manufacturer StreetLINAC HOUSE FLEMING WAY
Manufacturer CityCRAWLEY, RH109RR
Manufacturer CountryUK
Manufacturer PostalRH10 9RR
Manufacturer G1ELEKTA INSTRUMENT AB
Manufacturer StreetKUNGSTENSGATAN 18 PO BOX 7593
Manufacturer CitySTOCKHOLM, SE10393
Manufacturer CountrySW
Manufacturer Postal CodeSE103 93
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLEKSELL GAMMA KNIFE ICON
Generic NameSYSTEM, RADIATION THERAPY, RADIONUCLIDE
Product CodeIWB
Date Received2017-03-30
Model Number11.1
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerELEKTA INSTRUMENT AB
Manufacturer AddressKUNGSTENSGATAN 18 PO BOX 7593 STOCKHOLM, SE10393 SW SE103 93


Patients

Patient NumberTreatmentOutcomeDate
10 2017-03-30

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.