DIMENSION VISTA? FERRITIN FLEX? REAGENT CARTRIDGE K6440 (SMN 10445136)

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-03-30 for DIMENSION VISTA? FERRITIN FLEX? REAGENT CARTRIDGE K6440 (SMN 10445136) manufactured by Siemens Healthcare Diagnostics Inc - Glasgow.

Event Text Entries

[71335073] Siemens healthcare diagnostics has determined that dimension troponin (tni) and dimension vista digoxin (digxn), estradiol (e2), ferritin (ferr), prolactin (prl), and thyroid stimulating hormone (tsh) have incorrect units for botin in the instructions for use and incorrectly state the level at which biotin does not interfere. In the cases of tni, e2, ferr, prl, and tsh, there is a significant interference at the levels incorrectly stated in the current ifus. Since the biotin si units are listed as umol/l instead of nmol/l the si units are incorrect by a factor of 1000 higher than the level at which biotin does not interfere, e2, a competitive assay, exhibits a positive bias while the other methods (sandwich assays) exhibit a negative bias. Digxn si units for biotin are incorrect by a factor of 100 lower than the level at which biotin does not interfere, in the current ifu. Siemens issued an urgent medical device correction (udmc), dated march 29, 2017, communication (b)(4), to all us accounts and an urgent field safety notice (ufsn), dated march 2017, communication (b)(4), to all outside us accounts who had been shipped the impacted products. Customers were instructed that the non-interfering table for biotin in the letter supersedes the information related to this section in the current ifus for the assays listed until the ifus are updated.
Patient Sequence No: 1, Text Type: N, H10


[71335074] Customer reported that there is a typographical error in the dimension vista free thyroxine (ft4) for biotin in the non-interfering table of the instructions for use. Customer is comparing biotin syst? Me international d'unit? S [si units] for ft4 which are nmol/l to ferritin (ferr) and thyroid stimulating hormone (tsh) which are in umol/l. However, ft4 si units are correct and the biotin si units for ferr and tsh should be nmol/l. There are no reports of patient intervention or adverse health consequences due to the incorrect si units for biotin in ferr ifu.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2517506-2017-00299
MDR Report Key6445792
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2017-03-30
Date of Report2017-03-21
Date of Event2016-09-27
Date Added to Maude2017-03-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPH.D EDWARD SZYMANSKI
Manufacturer StreetGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE PO BOX 6101
Manufacturer CityNEWARK DE 197146101
Manufacturer CountryUS
Manufacturer Postal197146101
Manufacturer Phone3026317672
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS, INC - GLASGOW
Manufacturer StreetGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE PO BOX 6101
Manufacturer CityNEWARK DE 197146101
Manufacturer CountryUS
Manufacturer Postal Code197146101
Single Use3
Remedial ActionRC
Previous Use Code3
Removal Correction Number2517506-03/30/2017-003-C
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameDIMENSION VISTA? FERRITIN FLEX? REAGENT CARTRIDGE
Generic NameDIMENSION VISTA? FERR
Product CodeCEC
Date Received2017-03-30
Catalog NumberK6440 (SMN 10445136)
Lot NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No0
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Manufacturer AddressGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE PO BOX 6101 NEWARK DE 197146101 US 197146101

Device Sequence Number: 1

Brand NameDIMENSION VISTA? FERRITIN FLEX? REAGENT CARTRIDGE
Generic NameDIMENSION VISTA? FERR
Product CodeJMG
Date Received2017-03-30
Catalog NumberK6440 (SMN 10445136)
Lot NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Manufacturer AddressGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE PO BOX 6101 NEWARK DE 197146101 US 197146101


Patients

Patient NumberTreatmentOutcomeDate
10 2017-03-30

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.