CELLEBRITY? M00516151 1615

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2017-03-30 for CELLEBRITY? M00516151 1615 manufactured by Boston Scientific - Spencer.

Event Text Entries

[71276394] (b)(4). Investigation results: visual evaluation of the returned device revealed that the sheath was separated and torn at the proximal end. It was found that the catheter was kinked in multiple locations throughout. Functional analysis showed that the device handle moved smoothly in both directions, however, the brush failed to extend due to sheath was separated. Due to anatomical/procedural factors encountered during the procedure; performance of the device was limited. Therefore, the most probable root cause classification for the reported failure is operational context. A review of the device history record (dhr) was performed and no deviations were found. A search of the complaint database revealed that no other complaints exist for the specified lot.
Patient Sequence No: 1, Text Type: N, H10


[71276395] It was reported to boston scientific corporation that a cellebrity cytology brush was used during a procedure performed on (b)(6) 2017. According to the complainant, during the procedure, the brush failed to extend. The procedure was completed with another cellebrity brush. There were no known patient complications reported as a result of this event. This event has been deemed a reportable event based on the investigation results; sheath was separated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2017-00812
MDR Report Key6446604
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2017-03-30
Date of Report2017-03-09
Date of Event2017-02-13
Date Mfgr Received2017-03-09
Device Manufacturer Date2016-12-09
Date Added to Maude2017-03-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEMP. NANCY CUTINO
Manufacturer Street100 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834000
Manufacturer G1BOSTON SCIENTIFIC - SPENCER
Manufacturer Street780 BROOKSIDE DRIVE
Manufacturer CitySPENCER IN 47460
Manufacturer CountryUS
Manufacturer Postal Code47460
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCELLEBRITY?
Generic NameBRUSH, BIOPSY, BRONCHOSCOPE (NON-RIGID)
Product CodeBTG
Date Received2017-03-30
Returned To Mfg2017-02-23
Model NumberM00516151
Catalog Number1615
Lot Number20048635
Device Expiration Date2018-12-31
OperatorNURSE
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC - SPENCER
Manufacturer Address780 BROOKSIDE DRIVE SPENCER IN 47460 US 47460


Patients

Patient NumberTreatmentOutcomeDate
10 2017-03-30

© 2024 FDA.report