Event Text Entries

[71493604] Investigation including root cause analysis is in progress. A supplemental mdr will be filed as necessary in accordance with 21 cfr 803. 56 when additional reportable information becomes available. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[71493605] A healthcare professional reported that the laser power was poor. The procedure was completed with no harm to the patient.
Patient Sequence No: 1, Text Type: D, B5


[76891940] The system was examined and the reported event was not replicated. The company representative cleaned the optics block of the slit lamp and aligned the aiming be am to address the issue. The slip lamp was manufactured on march 31, 2010. Based on qa assessment, the system met specifications at the time of release. The system was tested and found to meet product specifications. The root cause of the reported event can be attributed to dust and debris accumulating on the slit lamp optics over time. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2028159-2017-01357
MDR Report Key6446971
Date Received2017-03-30
Date of Report2017-06-07
Date of Event2017-03-09
Date Mfgr Received2017-05-16
Date Added to Maude2017-03-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. RITA LOPEZ
Manufacturer Street6201 SOUTH FREEWAY MAIL STOP AB2-6
Manufacturer CityFORT WORTH TX 76134
Manufacturer CountryUS
Manufacturer Postal76134
Manufacturer Phone8175514846
Manufacturer G1ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
Manufacturer Street15800 ALTON PARKWAY
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal Code92618
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameSLIT LAMP
Generic NameBIOMICROSCOPE, SLIT-LAMP, AC-POWERED
Product CodeHJO
Date Received2017-03-30
Model NumberNA
Catalog Number8065740982
Lot NumberASKU
ID Number1.04_BLD_0043
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
Manufacturer Address15800 ALTON PARKWAY IRVINE CA 92618 US 92618


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-03-30