AMPLICHIP CYP450 TEST * 04591402190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,07 report with the FDA on 2005-09-30 for AMPLICHIP CYP450 TEST * 04591402190 manufactured by Roche Molecular Systems.

Event Text Entries

[428085] The customer was using the rms amplichip cyp450 kit. Out of approximately 83 samples of known genotype, one was reported as discrepant when compared to three other genotyping platforms. The rms amplichip generated a genotype result of 2d6 *1/*41 and the other platforms generated a 2d6 *3/*4 genotype. Genotype 2d6 *1/*41 is associated with a phenotypic expression of an extensive metabolizer and genotype 2d6 *3/*4 with a phenotypic expression of a poor metabolizer. These results are part of an evaluation study and therefore had no impact on pt treatment. Treatment was neither recommended nor changed based on these results.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2243471-2005-00024
MDR Report Key645640
Report Source05,07
Date Received2005-09-30
Date of Report2005-08-30
Date Mfgr Received2005-09-01
Date Added to Maude2005-11-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactROBERT PIGOZZI
Manufacturer Street1080 US HIGHWAY 202
Manufacturer CityBRANCHBURG NJ 08876
Manufacturer CountryUS
Manufacturer Postal08876
Manufacturer Phone9257308272
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAMPLICHIP CYP450 TEST
Generic NameMICRO ARRAY, IN VITRO DIAGNOSTIC
Product CodeNTI
Date Received2005-09-30
Model Number*
Catalog Number04591402190
Lot NumberG01491
ID NumberNA
Device Expiration Date2005-08-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key635147
ManufacturerROCHE MOLECULAR SYSTEMS
Manufacturer Address* BRANCHBURG NJ * US
Baseline Brand NameAMPLICHIP CYP450 TEST
Baseline Generic NameMICRO ARRAY, IN VITRO DIAGNOSTIC
Baseline Model No*
Baseline Catalog No04591402190
Baseline IDNA


Patients

Patient NumberTreatmentOutcomeDate
10 2005-09-30

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.